Voyager DP-263-S Guia do Utilizador Página 1

Consulte online ou descarregue Guia do Utilizador para Sistemas de porteiro eletrónico Voyager DP-263-S. Voyager DP-263-S User`s guide Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
Users Guide
Version 5 Series Software
PerSeptive Biosystems, Inc.
500 Old Connecticut Path
Framingham, MA 01701 USA
A subsidiary of PE Corporation
Part Number V900112-02, Rev. A
August 1999
Voyager
Biospectrometry
Workstation
with Delayed Extraction Technology
Vista de página 0
1 2 3 4 5 6 ... 611 612

Resumo do Conteúdo

Página 1 - Workstation

User’s GuideVersion 5 Series SoftwarePerSeptive Biosystems, Inc.500 Old Connecticut PathFramingham, MA 01701 USAA subsidiary of PE CorporationPart Nu

Página 2

Table of Contentsx PerSeptive BiosystemsChapter 7 PSD Analysis7.1 Overview of PSD Analysis...

Página 3

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-22 PerSeptive Biosystems2Video monitorsFigure 2-12 and Figure 2-13 show the re

Página 4

Connecting the Voyager-DE STR WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 2-232NOTE: In the US only, you can plug the video monitor

Página 5 - Chapter 3 Preparing Samples

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-24 PerSeptive Biosystems22.5 Installing SoftwareAll necessary software is ins

Página 6

Installing SoftwareVoyager™ Biospectrometry™ Workstation User’s Guide 2-2522.5.1 Installing the Voyager SoftwareInstallingTo install the Voyager soft

Página 7 - ... 5-39

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-26 PerSeptive Biosystems2• User Guides—Includes PDF (portable document format)

Página 8

Installing SoftwareVoyager™ Biospectrometry™ Workstation User’s Guide 2-2722.5.2 Starting the SoftwareStartingInstrumentControl PanelTo start the Voy

Página 9 - ... 6-60

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-28 PerSeptive Biosystems2StartingSequence ControlPanelTo start the Sequence Co

Página 10 - Chapter 7 PSD Analysis

Installing SoftwareVoyager™ Biospectrometry™ Workstation User’s Guide 2-2922.5.3 Exiting the SoftwareCAUTIONIf you are using the Voyager Workstation

Página 11 - Table of Contents

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-30 PerSeptive Biosystems22.6 Hardware ConfigurationCAUTIONDo not alter the Ha

Página 12

Hardware ConfigurationVoyager™ Biospectrometry™ Workstation User’s Guide 2-312.Figure 2-16 Vacuum Configuration3. Check the following values as neede

Página 13 - Instrument Safety

Table of ContentsVoyager Biospectrometry Workstation User’s Guide xiAppendix A Specifications...

Página 14

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-32 PerSeptive Biosystems2• Source Chamber (BA1) Wait Time—Time (seconds) that

Página 15 - Electrical Symbols

Hardware ConfigurationVoyager™ Biospectrometry™ Workstation User’s Guide 2-332High voltageconfigurationTo check high voltage configuration:1. In the I

Página 16 - Non-electrical

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-34 PerSeptive Biosystems2• Idle Power Off/Idle Time—When enabled, the number o

Página 17

Hardware ConfigurationVoyager™ Biospectrometry™ Workstation User’s Guide 2-3523. Check the following values as needed:• Flight Length to Deflector (Re

Página 18 - Symboles non

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-36 PerSeptive Biosystems23. Check the following fields as needed:• Instrument

Página 19 - Workstation User’s Guide xix

Hardware ConfigurationVoyager™ Biospectrometry™ Workstation User’s Guide 2-372Figure 2-20 Laser Configuration3. Check the following fields as needed:

Página 20

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-38 PerSeptive Biosystems2DigitizerconfigurationTo check the configuration of t

Página 21 - Workstation User’s Guide xxi

Aligning the Sample PlateVoyager™ Biospectrometry™ Workstation User’s Guide 2-3922.7 Aligning the Sample PlateIn this sectionThis section describes:•

Página 22

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-40 PerSeptive Biosystems2.PLT files andmultiplealignmentsA .PLT file defines t

Página 23 - Safety and EMC

Aligning the Sample PlateVoyager™ Biospectrometry™ Workstation User’s Guide 2-412Overview of videomonitor displayThe following examples show how the s

Página 24 - Sécurité

Table of Contentsxii PerSeptive Biosystems

Página 25 - Laser Safety

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-42 PerSeptive Biosystems2What you needTo align the sample plate you need the f

Página 26

Aligning the Sample PlateVoyager™ Biospectrometry™ Workstation User’s Guide 2-432Using the controlstickTo align a sample plate, you must use the contr

Página 27 - How to Use This Guide

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-44 PerSeptive Biosystems2Figure 2-26 Sample Plate Alignment Wizard3. Click Ne

Página 28

Aligning the Sample PlateVoyager™ Biospectrometry™ Workstation User’s Guide 2-452The software calculates the alignment and uses the settings to ensure

Página 29

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-46 PerSeptive Biosystems22.8 Startup and ShutdownIn this sectionThis section

Página 30

Startup and ShutdownVoyager™ Biospectrometry™ Workstation User’s Guide 2-472If your system includes a LeCroy digitizer, wait approximately one minute

Página 31 - Workstations

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-48 PerSeptive Biosystems2InitializingThe hardware is automatically initialized

Página 32 - System Overview

Startup and ShutdownVoyager™ Biospectrometry™ Workstation User’s Guide 2-492Powering downthe massspectrometerIf you need to perform maintenance on int

Página 33

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-50 PerSeptive Biosystems22.9 Checking System StatusDisplay the System Status

Página 34 - Features

Checking System StatusVoyager™ Biospectrometry™ Workstation User’s Guide 2-512Table 2-2 System Status ParametersParameter DescriptionSystem StatusIn

Página 35 - 1.2 Voyager-DE STR

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xiii1Safety and Compliance InformationIn this sectionThis section

Página 36

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-52 PerSeptive Biosystems2

Página 37 - Technology Overview

3ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 3-13 Preparing SamplesThis chapter contains the following sections:3.1 Preparing Samples

Página 38 - Ionization

Chapter 3 Preparing Samples3-2 PerSeptive Biosystems33.1 Preparing Samples NOTE: Sample preparation is critical to obtain good results in MALDI-T

Página 39 - Workstation User’s Guide 1-9

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-333.1.1 Selecting a MatrixPurpose of matrixIn MALDI-TOF, the matrix plays a key

Página 40 - MALDI-TOF

Chapter 3 Preparing Samples3-4 PerSeptive Biosystems3MixturesIf you are examining a sample mixture, it may be necessary to prepare the mixture wit

Página 41 - Technology

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-53NOTE: HPLC-grade water may vary in salt concentration and may produce adduct i

Página 42 - 1-12 PerSeptive Biosystems

Chapter 3 Preparing Samples3-6 PerSeptive Biosystems35. Cap the tube and vortex thoroughly for approximately 1 minute or until dissolved. You can

Página 43 - Extraction (Linear Mode)

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-73Sinapinic acidUse sinapinic acid for peptides and proteins >10,000 Da.Matri

Página 44 - Extraction

Chapter 3 Preparing Samples3-8 PerSeptive Biosystems3CHCAUseα-cyano-4-hydroxycinnamic acid (CHCA) for:• Dried drop application—Peptides/proteins &

Página 45 - Velocity focusing

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-93CHCA for thin layerapplicationUse for peptides <3,000 Da.Matrix concentrati

Página 46 - —Detection

Safety and Compliance Informationxiv PerSeptive Biosystems1A Caution provides information to avoid damage to the system or loss of data and appears as

Página 47 - PB100465

Chapter 3 Preparing Samples3-10 PerSeptive Biosystems3THAPUse THAP for small oligonucleotides <3,500 Da, acidic carbohydrates, acidic glycopept

Página 48 - PB100466

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-1133-HPAUse 3-HPA for large oligonucleotides >3,500 Da.Matrix concentration 5

Página 49

Chapter 3 Preparing Samples3-12 PerSeptive Biosystems3DHBUse DHB for:• Neutral carbohydrates• Small moleculesDHB for neutralcarbohydratesMatrix co

Página 50 - 1.5.2 Mass Spectrometer

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-133DHB for smallmoleculesMatrix concentration 10 mg/mlFinal sample concentration

Página 51

Chapter 3 Preparing Samples3-14 PerSeptive Biosystems3DHBsUse DHBs for peptides and proteins >10,000 Da, and glycosylated proteins.Matrix conce

Página 52 - Parts of the mass

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-153Synthetic polymermatricesChoice of matrix depends on the sample you are analy

Página 53

Chapter 3 Preparing Samples3-16 PerSeptive Biosystems3You can load polymer samples on sample plates in two ways:• Thin layer polymer method (yield

Página 54 - 1.5.3 Vacuum System

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-1733.1.4 Preparing SampleIn this sectionThis section includes:• Sample concentr

Página 55

Chapter 3 Preparing Samples3-18 PerSeptive Biosystems3Note the following:• When diluting sample, keep in mind that you will be further diluting wh

Página 56 - Vacuum gauges

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-1933.1.5 Sample CleanupCleaningtechniquesUse the following techniques to clean

Página 57

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xv1Safety symbolsThe following symbols may be displayed on the sys

Página 58

Chapter 3 Preparing Samples3-20 PerSeptive Biosystems33.1.5.1 WashingWhen to useUse this technique when you know the solubility of the contaminan

Página 59 - 1.5.4 Vacuum Gauge Panel

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-2133.1.5.2 Drop Dialysis (Floating Membrane)When to useUse this technique on po

Página 60 - 1-30 PerSeptive Biosystems

Chapter 3 Preparing Samples3-22 PerSeptive Biosystems3NOTE: Sample also passes through the membrane during dialysis, particularly low mass samples

Página 61 - 1.5.5 Computer Components

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-2333.1.5.3 Cation ExchangeWhen to useUse this technique for non-polar proteins

Página 62 - 1.6.1 System Components

Chapter 3 Preparing Samples3-24 PerSeptive Biosystems33.1.5.4 ZipTips® When to useUse this technique for peptides, proteins, and oligonucleotides

Página 63

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-253ProcedureTo clean samples with ZipTips:1. Wash a C18 ZipTip in the following

Página 64 - 1.6.2 Mass Spectrometer

Chapter 3 Preparing Samples3-26 PerSeptive Biosystems33.1.6 Mixing Sample and Matrix (Dried Droplet Application)When to useUse dried droplet appl

Página 65

Preparing SamplesVoyager™ Biospectrometry™ Workstation User’s Guide 3-273Mixing sampleand matrix on thesample plateMix sample and matrix directly on t

Página 66 - 4113, and 4116 and later

Chapter 3 Preparing Samples3-28 PerSeptive Biosystems33.2 Loading Samples on Sample PlatesIn this sectionThis section describes:•Overview• Loadin

Página 67

Loading Samples on Sample PlatesVoyager™ Biospectrometry™ Workstation User’s Guide 3-293Types of sampleplatesThree types of 100-position reusable samp

Página 68 - 1.6.3 Vacuum System

Safety and Compliance Informationxvi PerSeptive Biosystems1Non-electricalSymbolsThe following is an illustrated glossary of non-electrical safety aler

Página 69 - (Top View)

Chapter 3 Preparing Samples3-30 PerSeptive Biosystems3Handling sampleplatesTo prevent contamination of your analysis:• Start with a clean sample p

Página 70 - 1.6.4 Vacuum Gauge Panel

Loading Samples on Sample PlatesVoyager™ Biospectrometry™ Workstation User’s Guide 3-3133.2.2 Loading Samples (Dried Droplet Application)NOTE: If you

Página 71

Chapter 3 Preparing Samples3-32 PerSeptive Biosystems3Do not touch the tip of the pipette to the sample spot. Make sure the sample is evenly appli

Página 72 - 1.6.5 Front Panel Indicators

Loading Samples on Sample PlatesVoyager™ Biospectrometry™ Workstation User’s Guide 3-3333.2.3 Loading Samples (Thin Layer Application)When to useUse

Página 73 - 1.6.6 Computer Components

Chapter 3 Preparing Samples3-34 PerSeptive Biosystems33.2.4 Examining Crystals on Sample PlatesOverviewIf you are using polished blank or laser e

Página 74 - Software

Loading Samples on Sample PlatesVoyager™ Biospectrometry™ Workstation User’s Guide 3-353You can acquire data from a well that does not have an ideal c

Página 75 - 1.7 Software Overview

Chapter 3 Preparing Samples3-36 PerSeptive Biosystems33.3 Cleaning Sample PlatesIn this sectionThis section describes procedures for cleaning:•Te

Página 76 - 1-46 PerSeptive Biosystems

Cleaning Sample PlatesVoyager™ Biospectrometry™ Workstation User’s Guide 3-3736. Examine the plate. If you see any sample or matrix residue, oil, or f

Página 77

Chapter 3 Preparing Samples3-38 PerSeptive Biosystems3CAUTIONDo not leave the sample plate in detergent for longer than 10 minutes. Longer exposur

Página 78 - (Data Explorer)

Loading Sample Plates in the Mass SpectrometerVoyager™ Biospectrometry™ Workstation User’s Guide 3-3933.4 Loading Sample Plates in the Mass Spectrome

Página 79 - Biospectrometry™

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xvii1Symboles desalertes desécuritéLes symboles suivants peuvent ê

Página 80 - Voyager-DE PRO Workstations

Chapter 3 Preparing Samples3-40 PerSeptive Biosystems3Table 3-1 Sample Plate InformationParameter DescriptionPlate ID Unique identifier (up to 3

Página 81 - Workstation User’s Guide 2-3

Loading Sample Plates in the Mass SpectrometerVoyager™ Biospectrometry™ Workstation User’s Guide 3-4133.4.1 Ejecting the Sample HolderIn the Voyager

Página 82 - PB100507

Chapter 3 Preparing Samples3-42 PerSeptive Biosystems33.4.2 Loading Sample PlatesThis section describes loading sample plates in:• Voyager-DE and

Página 83

Loading Sample Plates in the Mass SpectrometerVoyager™ Biospectrometry™ Workstation User’s Guide 3-433Figure 3-5 Loading the Sample Plate in a Voyage

Página 84 - 2-6 PerSeptive Biosystems

Chapter 3 Preparing Samples3-44 PerSeptive Biosystems35. Select a Plate ID. The .PLT file and alignment information associated with the Plate ID a

Página 85

Loading Sample Plates in the Mass SpectrometerVoyager™ Biospectrometry™ Workstation User’s Guide 3-453Figure 3-7 Loading the Sample Plate in a Voyage

Página 86 - 2-8 PerSeptive Biosystems

Chapter 3 Preparing Samples3-46 PerSeptive Biosystems35. Select a Plate ID. The .PLT file and alignment information associated with the Plate ID a

Página 87 - PB100776

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-4733.5 Sample Plate Types SupportedIn addition to 100-position reusa

Página 88 - PB100784

Chapter 3 Preparing Samples3-48 PerSeptive Biosystems33.5.1 Sample Plates Types and ApplicationsThe Voyager Instrument Control software supports

Página 89 - 2.3.2 Connecting the

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-493NOTE: For additional information on available sample plates and or

Página 90 - 2.3.3 Connecting the

Safety and Compliance Informationxviii PerSeptive Biosystems1Symboles nonélectriquesLe tableau suivant donne la signification des symboles d’alertes d

Página 91 - LSA1000 LeCroy Digitizer

Chapter 3 Preparing Samples3-50 PerSeptive Biosystems33.5.2 Editable-Configuration Plate (.PLT) Types Provided with the SystemSelecting platetype

Página 92 - 2-14 PerSeptive Biosystems

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-513.PLT filesprovidedThe editable-configuration plate types provided

Página 93 - PB100770

Chapter 3 Preparing Samples3-52 PerSeptive Biosystems33.5.3 Guidelines forDefining Custom Plate TypesYou can define custom plate types of unlimit

Página 94 - 2-16 PerSeptive Biosystems

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-533A .PLT file describes each position on the plate as discussed in T

Página 95

Chapter 3 Preparing Samples3-54 PerSeptive Biosystems3X and Y Coordinates of the position center in reference to the lower left corner of the samp

Página 96 - NOTE: There are two 15-pin

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-553Width Use if you are customizing a .PLT file. Defines the width of

Página 97 - Voyager-DE STR Workstation

Chapter 3 Preparing Samples3-56 PerSeptive Biosystems3Figure 3-11 illustrates the coordinates for the four corner positions that define the edges

Página 98

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-573Guidelines forcreating .PLT filesNote the following:• The diameter

Página 99

Chapter 3 Preparing Samples3-58 PerSeptive Biosystems3NOTE: You can correct for systematic errors introduced by a sample preparation device by ali

Página 100 - Video monitors

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-593Tolerances andnon-systematicerrorsWhen creating search pattern fil

Página 101 - Biospectrometry

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xix1Before operatingthis instrumentEnsure that anyone involved wit

Página 102 - 2.5 Installing Software

Chapter 3 Preparing Samples3-60 PerSeptive Biosystems3Figure 3-12 Area Available for AnalysisHow to create a.PLT fileNOTE: Display the Sample Vie

Página 103 - Installing Software

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-613Figure 3-13 shows the Plate View displaying the logical coordinate

Página 104 - Accessing the

Chapter 3 Preparing Samples3-62 PerSeptive Biosystems38. On the second line, type in absolute coordinates for the first position. Separate absolut

Página 105 - 2.5.2 Starting the Software

Sample Plate Types SupportedVoyager™ Biospectrometry™ Workstation User’s Guide 3-6333.5.4 Adjusting the Laser Position for a Custom .PLT File After y

Página 106 - 2-28 PerSeptive Biosystems

Chapter 3 Preparing Samples3-64 PerSeptive Biosystems3

Página 107 - 2.5.3 Exiting the Software

4ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 4-14 Voyager Instrument Control Panel BasicsThis chapter contains the following sections:4

Página 108 - 2.6 Hardware Configuration

Chapter 4 Voyager Instrument Control Panel Basics4-2 PerSeptive Biosystems44.1 Instrument Control PanelThe Voyager Instrument Control Panel allow

Página 109 - Hardware Configuration

Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 4-34Figure 4-1 Instrument Control PanelToolbarThe toolbar contains buttons

Página 110 - 2-32 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-4 PerSeptive Biosystems4Control pagesThe Instrument Control Panel contains five control pages:•

Página 111 - High voltage

Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 4-54Spectrum windowThe Spectrum window provides a display of data. The data

Página 112 - Timed ion

NOTICEPerSeptive Biosystems, Inc. supplies or recommends certain configurations of computer hardware, software, and peripherals for use with its instr

Página 113 - Instrument

Safety and Compliance Informationxx PerSeptive Biosystems1General WarningsWARNINGFIRE HAZARD. Using a fuse of the wrong type or rating can cause a fir

Página 114 - 2-36 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-6 PerSeptive Biosystems4DisplayingThe Output window is automatically displayed when you:• Acqui

Página 115

Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 4-74Automatic controlmodeTo select Automatic Control mode:1. Open the Instr

Página 116 - Digitizer

Chapter 4 Voyager Instrument Control Panel Basics4-8 PerSeptive Biosystems44.2 Using the Control PagesThe Instrument Control Panel allows you to

Página 117 - When to align

Using the Control PagesVoyager™ Biospectrometry™ Workstation User’s Guide 4-94Types of pagecontrolYou can select between two types of page control for

Página 118 - 2-40 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-10 PerSeptive Biosystems44.3 Using the Spectrum WindowThis section includes:• Adjusting the di

Página 119 - Alignment

Using the Spectrum WindowVoyager™ Biospectrometry™ Workstation User’s Guide 4-1144. Set From and To values for the display range (m/z units) and click

Página 120 - 2-42 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-12 PerSeptive Biosystems47. Set as needed:8. Click OK.Parameter DescriptionScaling ModeDisplay

Página 121 - Aligning

Using the Spectrum WindowVoyager™ Biospectrometry™ Workstation User’s Guide 4-1344.3.2 Zooming on TracesZooming andunzoomingYou can expand (zoom) an

Página 122 - 2-44 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-14 PerSeptive Biosystems4Types of tracesThe Spectrum window can contain two types of traces (se

Página 123 - Aligning the Sample Plate

Using the Spectrum WindowVoyager™ Biospectrometry™ Workstation User’s Guide 4-154Hint: Resize the window to view all added traces.Figure 4-5 Adding T

Página 124 - 2.8 Startup and Shutdown

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xxi1WARNINGELECTRICAL SHOCK HAZARD. Severe electrical shock can re

Página 125 - Startup and Shutdown

Chapter 4 Voyager Instrument Control Panel Basics4-16 PerSeptive Biosystems44.3.4 Annotating TracesThis section describes:• Two ways to annotate•

Página 126 - 2-48 PerSeptive Biosystems

Using the Spectrum WindowVoyager™ Biospectrometry™ Workstation User’s Guide 4-174Hint: To move the text, left-click and hold on the text, then drag to

Página 127 - Powering down

Chapter 4 Voyager Instrument Control Panel Basics4-18 PerSeptive Biosystems44.3.5 Previewing and Printing TracesThis section includes:• Previewin

Página 128 - 2.9 Checking System Status

Using the Spectrum WindowVoyager™ Biospectrometry™ Workstation User’s Guide 4-194NOTE: If you set Landscape printing orientation within Instrument Con

Página 129 - Checking System Status

Chapter 4 Voyager Instrument Control Panel Basics4-20 PerSeptive Biosystems4Dedicating aprinter tolandscapeorientationTo dedicate the printer to l

Página 130 - 2-52 PerSeptive Biosystems

Customizing the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 4-2144.4 Customizing the Instrument Control PanelUndocking

Página 131 - 3 Preparing Samples

Chapter 4 Voyager Instrument Control Panel Basics4-22 PerSeptive Biosystems4Accessinggraphic optionsTo access the graphic options:1. Display the t

Página 132 - 3.1 Preparing Samples

Customizing the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 4-234Figure 4-6 Graph and Plot Options Dialog BoxFor addit

Página 133 - 3.1.1 Selecting a Matrix

Chapter 4 Voyager Instrument Control Panel Basics4-24 PerSeptive Biosystems44.5 Controlling the WorkstationThis section includes:• Using toolbar

Página 134 - 3.1.2 Preparing Matrix

Controlling the WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 4-254Turninghigh voltage onand offClick in the toolbar to turn the hig

Página 135 - Workstation User’s Guide 3-5

Safety and Compliance Informationxxii PerSeptive Biosystems1WARNINGPHYSICAL INJURY HAZARD. Use the Voyager Biospectrometry Workstation only as specifi

Página 136 - 3.1.3 Matrix Information

Chapter 4 Voyager Instrument Control Panel Basics4-26 PerSeptive Biosystems4AccumulatingspectraClick in the toolbar to manually accumulate spect

Página 137 - Workstation User’s Guide 3-7

Controlling the WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 4-2744.5.2 Adjusting Laser Intensity and Selecting Sample PositionThis

Página 138 - 3-8 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-28 PerSeptive Biosystems4The Manual Laser/Sample Position control page allows you to:• Manually

Página 139 - Workstation User’s Guide 3-9

Controlling the WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 4-294Selecting theactive sampleposition in PlateviewThe active sample po

Página 140 - 3-10 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-30 PerSeptive Biosystems4You can use these coordinates when you create a search pattern file. F

Página 141

Controlling the WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 4-314Hint: If you right-click on a position, you can change between Samp

Página 142 - DHB for neutral

Chapter 4 Voyager Instrument Control Panel Basics4-32 PerSeptive Biosystems44.6 Sequence Control PanelSequence ControlPanelThe Voyager Sequence C

Página 143 - DHB for small

How the Instrument and Sequence Control Panels InteractVoyager™ Biospectrometry™ Workstation User’s Guide 4-334When you start the Sequence Control Pan

Página 144 - 3-14 PerSeptive Biosystems

Chapter 4 Voyager Instrument Control Panel Basics4-34 PerSeptive Biosystems4If you start the Instrument Control Panel using the Sequence Control P

Página 145 - Synthetic polymer

How the Instrument and Sequence Control Panels InteractVoyager™ Biospectrometry™ Workstation User’s Guide 4-354Keeping bothcontrol panelsopenYou can k

Página 146 - 3-16 PerSeptive Biosystems

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xxiii1Safety and EMC (Electromagnetic Compliance) StandardsUS Safe

Página 147 - 3.1.4 Preparing Sample

Chapter 4 Voyager Instrument Control Panel Basics4-36 PerSeptive Biosystems4

Página 148 - Internal standards

5ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 5-1]75 OptimizingInstrument SettingsThis chapter contains the following sections:5.1 Load

Página 149 - 3.1.5 Sample Cleanup

Chapter 5 Optimizing Instrument Settings5-2 PerSeptive Biosystems55.1 Loading, Modifying, and Saving Instrument SettingsThis section includes:• U

Página 150 - 3.1.5.1 Washing

Loading, Modifying, and Saving Instrument SettingsVoyager™ Biospectrometry™ Workstation User’s Guide 5-35NOTE: Data storage parameters are not stored

Página 151 - Procedure

Chapter 5 Optimizing Instrument Settings5-4 PerSeptive Biosystems5List of .BIC filesTable 5-1 through Table 5-3 list the standard .BIC files provi

Página 152 - 3-22 PerSeptive Biosystems

Loading, Modifying, and Saving Instrument SettingsVoyager™ Biospectrometry™ Workstation User’s Guide 5-55Table 5-2 Reflector Mode .BIC Files.BIC Fil

Página 153 - 3.1.5.3 Cation Exchange

Chapter 5 Optimizing Instrument Settings5-6 PerSeptive Biosystems5NOTE: For CID applications, use Substance_P_Precursor.BIC method with the approp

Página 154 - 3.1.5.4 ZipTips

Loading, Modifying, and Saving Instrument SettingsVoyager™ Biospectrometry™ Workstation User’s Guide 5-755.1.3 Opening and Viewing Instrument Setting

Página 155

Chapter 5 Optimizing Instrument Settings5-8 PerSeptive Biosystems5Opening from .DATTo open an instrument settings file from a .DAT file, select .D

Página 156 - (Dried Droplet Application)

Loading, Modifying, and Saving Instrument SettingsVoyager™ Biospectrometry™ Workstation User’s Guide 5-95Modifying forManual ControlmodeTo modify the

Página 157 - Mixing sample

Safety and Compliance Informationxxiv PerSeptive Biosystems1Canadian Safetyand EMCStandardsSafetyThis instrument has been tested to and complies with

Página 158 - 3.2.1 Overview

Chapter 5 Optimizing Instrument Settings5-10 PerSeptive Biosystems53. Select Manual Control mode. For parameter descriptions, see Section 5.2.1, I

Página 159 - Types of sample

Loading, Modifying, and Saving Instrument SettingsVoyager™ Biospectrometry™ Workstation User’s Guide 5-1155.1.5 Saving and Printing Instrument Settin

Página 160 - 3-30 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-12 PerSeptive Biosystems5Saving .BIC filesfor differentmodesWhen saving .BIC files for use in different

Página 161 - 3.2.2 Loading Samples

Loading, Modifying, and Saving Instrument SettingsVoyager™ Biospectrometry™ Workstation User’s Guide 5-1355.1.6 Setting Instrument Settings Files to

Página 162 - 3-32 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-14 PerSeptive Biosystems55.2 Instrument Settings Parameter DescriptionsThis section describes the param

Página 163 - (Thin Layer Application)

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-1555.2.1 Instrument Settings PageSelect Instrument Set

Página 164 - PB100265

Chapter 5 Optimizing Instrument Settings5-16 PerSeptive Biosystems5Instrument settings parameters are described in Table 5-4.Table 5-4 Instrumen

Página 165 - PB100510

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-175VoltagesAccelerating Voltage Voltage applied to the

Página 166 - 3.3 Cleaning Sample Plates

Chapter 5 Optimizing Instrument Settings5-18 PerSeptive Biosystems5Guide Wire Voltage% Voltage applied to the beam guide wire. Overcomes the dispe

Página 167 - Cleaning Sample Plates

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-195Shots/Spectrum Determines the number of laser shots

Página 168 - 3-38 PerSeptive Biosystems

Safety and Compliance InformationVoyager™ Biospectrometry™ Workstation User’s Guide xxv1Laser SafetyLaserclassificationThe Voyager™ Biospectrometry™ W

Página 169 - 3.4 Loading Sample Plates

Chapter 5 Optimizing Instrument Settings5-20 PerSeptive Biosystems5Spectrum Acquisition (continued)Low Mass Gate (Da) Turns on the detector voltag

Página 170 - 3-40 PerSeptive Biosystems

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-215Calibration (continued)Default Enables default calib

Página 171

Chapter 5 Optimizing Instrument Settings5-22 PerSeptive Biosystems5Matrix influenceThe initial velocity is the average speed at which matrix ions

Página 172 - 3.4.2 Loading Sample Plates

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-235Modifying thematrix referencefileMatrix options are

Página 173 - 84 85 86

Chapter 5 Optimizing Instrument Settings5-24 PerSeptive Biosystems55.2.2 Mode/Digitizer Settings Dialog BoxClick Mode/Digitizer Settings in the I

Página 174 - 3-44 PerSeptive Biosystems

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-255Instrument ModeparametersClick the Instrument Mode t

Página 175 - PB100277

Chapter 5 Optimizing Instrument Settings5-26 PerSeptive Biosystems5Linear/ReflectorDigitizerparametersClick the Linear or Reflector Digitizer tab

Página 176 - 3-46 PerSeptive Biosystems

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-275 Table 5-7 Linear and Reflector Digitizer Paramete

Página 177 - Sample Plate Types Supported

Chapter 5 Optimizing Instrument Settings5-28 PerSeptive Biosystems5Vertical SettingsVertical Scale Specifies the input range of the digitizer in m

Página 178 - 3-48 PerSeptive Biosystems

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-295AdvancedparametersClick the Advanced tab to display

Página 179

Safety and Compliance Informationxxvi PerSeptive Biosystems1

Página 180 - 3-50 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-30 PerSeptive Biosystems5Table 5-8 Advanced ParametersParameter DescriptionMirror to Accelerating Volt

Página 181 - .PLT files

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-3155.2.3 Automatic Control Dialog BoxOn the Instrument

Página 182 - Defining Custom Plate Types

Chapter 5 Optimizing Instrument Settings5-32 PerSeptive Biosystems5For information on setting Instrument Settings for Automatic Control mode, see

Página 183

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-335Table 5-10 Automatic Control Parameters—Spectrum A

Página 184 - 3-54 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-34 PerSeptive Biosystems5Save Conditions(continued)• Save all spectra that pass acceptance criteria— Eac

Página 185

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-355Save Conditions(continued)• Save first spectrum to p

Página 186 - 3-56 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-36 PerSeptive Biosystems5Accumulation ConditionsThe following Accumulation conditions create one data fi

Página 187

Instrument Settings Parameter DescriptionsVoyager™ Biospectrometry™ Workstation User’s Guide 5-375Accumulation Conditions(continued)Accumulate all pas

Página 188 - 3-58 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-38 PerSeptive Biosystems5Table 5-11 Automatic Control Parameters—Sample PositioningParameter Descripti

Página 189

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-3955.3 Impact of Changing Instrument Settings P

Página 190 - 3-60 PerSeptive Biosystems

How to Use This GuideVoyager Biospectrometry Workstation User’s Guide xxvii1How to Use This GuidePurpose of thisguidePerSeptive Biosystems’ Voyager Bi

Página 191

Chapter 5 Optimizing Instrument Settings5-40 PerSeptive Biosystems5Grid Voltage% Linear/ ReflectorCritical parameter with optimum value for maximu

Página 192 - 3-62 PerSeptive Biosystems

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-4155.3.2 Understanding Grid Voltage%NOTE: You m

Página 193 - 3.5.4 Adjusting the Laser

Chapter 5 Optimizing Instrument Settings5-42 PerSeptive Biosystems5Potential gradientThe potential gradient in the ionization region (Figure 5-8)

Página 194 - 3-64 PerSeptive Biosystems

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-435For example (Figure 5-8), with a 25,000 V Acc

Página 195 - Panel Basics

Chapter 5 Optimizing Instrument Settings5-44 PerSeptive Biosystems55.3.3 Understanding Delay TimeDelay Time is the time in nanoseconds after the

Página 196 - 4.1 Instrument Control Panel

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-455Relationship toGrid Voltage%Delay Time and Gr

Página 197 - Workstation User’s Guide 4-3

Chapter 5 Optimizing Instrument Settings5-46 PerSeptive Biosystems55.3.4 Understanding Guide Wire Voltage%By applying voltage to the beam guide w

Página 198 - Voyager-DE STR system

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-475Reflector mode• To obtain maximum resolution

Página 199 - Output window

Chapter 5 Optimizing Instrument Settings5-48 PerSeptive Biosystems5Default Number ofData Points/BinSize settingsThe software uses the following de

Página 200 - Manual control

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-495Effects ofadjusting Bin sizeA smaller Bin siz

Página 201 - Workstation User’s Guide 4-7

How to Use This Guidexxviii PerSeptive Biosystems1Chapter 6, Acquiring Mass SpectraDescribes how to use the Voyager processing software, including lab

Página 202 - 4.2 Using the Control Pages

Chapter 5 Optimizing Instrument Settings5-50 PerSeptive Biosystems55.3.5.2 Vertical Digitizer SettingsThis section describes:• Vertical settings•

Página 203 - Types of page

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-515Figure 5-12 Effect of Vertical Scale in Digi

Página 204 - 4-10 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-52 PerSeptive Biosystems5When to increaseIncrease the Vertical Scale if signal goes offscale (Figure 5-1

Página 205 - Using the Spectrum Window

Impact of Changing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-535Input Bandwidth(not available withSignatecdig

Página 206 - 4-12 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-54 PerSeptive Biosystems55.4 Optimizing Instrument Settings ParametersIn this sectionThis section descr

Página 207 - 4.3.2 Zooming on Traces

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-5555.4.1 Optimization StrategyThis section gives a sugg

Página 208 - 4-14 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-56 PerSeptive Biosystems5StrategyTo optimize instrument settings, do the following:1. Open a .BIC file f

Página 209 - Removing traces

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-5755.4.2 Determining the Laser SettingThis section incl

Página 210 - 4.3.4 Annotating Traces

Chapter 5 Optimizing Instrument Settings5-58 PerSeptive Biosystems5Adjusting laserintensityTo efficiently optimize the laser setting, increase or

Página 211

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-595• If the spectrum you obtain is not acceptable, incre

Página 212 - 4-18 PerSeptive Biosystems

How to Use This GuideVoyager Biospectrometry Workstation User’s Guide xxix1RelateddocumentationThese related documents are shipped with your system:•V

Página 213 - 5. Click Print

Chapter 5 Optimizing Instrument Settings5-60 PerSeptive Biosystems5NOTE: Saturated signal in any region of the spectrum may suppress peaks in the

Página 214 - 4-20 PerSeptive Biosystems

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-6155.4.3 Optimizing ResolutionThis section includes:•Ov

Página 215 - 4.4 Customizing the

Chapter 5 Optimizing Instrument Settings5-62 PerSeptive Biosystems55.4.3.1 OverviewThis section includes:• Parameters to adjust•DE parameters • R

Página 216 - 4-22 PerSeptive Biosystems

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-635DE parametersImproved resolution in Delayed Extractio

Página 217

Chapter 5 Optimizing Instrument Settings5-64 PerSeptive Biosystems5Effects of massand matrixNote the following:• At a fixed Delay Time, higher mas

Página 218 - Instrument Menu Commands

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-6555.4.3.2 Acceptable Resolution in Delayed Extraction

Página 219 - Controlling the Workstation

Chapter 5 Optimizing Instrument Settings5-66 PerSeptive Biosystems5If you cannot obtain the resolution listed for the mass range of interest, cons

Página 220 - 4-26 PerSeptive Biosystems

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-6755.4.3.4 Optimizing Delay TimeThis section includes:•

Página 221 - Selecting Sample Position

Chapter 5 Optimizing Instrument Settings5-68 PerSeptive Biosystems5Optimizing DelayTimeNOTE: If you are analyzing a broad mass range and need opti

Página 222 - 4-28 PerSeptive Biosystems

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-6953. If the resolution improves by at least 20 percent

Página 223

Table of ContentsVoyager Biospectrometry Workstation User’s Guide iiiTable of ContentsSafety and Compliance Information ...

Página 224 - 4-30 PerSeptive Biosystems

How to Use This Guidexxx PerSeptive Biosystems1You also receive the appropriate manual for the digitizer included with your system:• Signatec Digitize

Página 225

Chapter 5 Optimizing Instrument Settings5-70 PerSeptive Biosystems5NOTE: Changing the Delay Time in increments smaller than 25 nsec may have no si

Página 226 - 4.6 Sequence Control Panel

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-7154. Continue increasing (or decreasing) the Delay Time

Página 227 - Control Panels Interact

Chapter 5 Optimizing Instrument Settings5-72 PerSeptive Biosystems55.4.3.5 Optimizing Grid Voltage%OverviewYou can use this procedure as an alter

Página 228 - 4-34 PerSeptive Biosystems

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-735Optimizing GridVoltage%To optimize Grid Voltage%:1. O

Página 229 - Sequence Control Panel

Chapter 5 Optimizing Instrument Settings5-74 PerSeptive Biosystems5If the resolution does not improve, decrease the Grid Voltage% (by 0.5 percent

Página 230 - 4-36 PerSeptive Biosystems

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-7555.4.4 Optimizing Signal-to-Noise RatioThis section i

Página 231 - Instrument Settings

Chapter 5 Optimizing Instrument Settings5-76 PerSeptive Biosystems55.4.4.1 OverviewYou can set the following parameters to optimize signal-to-noi

Página 232 - Saving Instrument Settings

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-7755.4.4.2 Setting Accelerating VoltageOverviewAccelera

Página 233 - 5.1.2 Standard Instrument

Chapter 5 Optimizing Instrument Settings5-78 PerSeptive Biosystems5A lower Accelerating Voltage setting does the following:• Provides more data po

Página 234 - List of .BIC files

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-7955.4.4.4 Setting Shots/SpectrumOverviewA higher numbe

Página 235 - Workstation User’s Guide 5-5

1ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 1-11 Introducing theVoyager™Biospectrometry™WorkstationsThis chapter contains the followin

Página 236 - 5-6 PerSeptive Biosystems

Chapter 5 Optimizing Instrument Settings5-80 PerSeptive Biosystems5When to useTurn on Low Mass Gate when:• Analyzing masses greater than 2,000–3,0

Página 237 - 5.1.3 Opening and Viewing

Optimizing Instrument Settings ParametersVoyager™ Biospectrometry™ Workstation User’s Guide 5-815Figure 5-20 Low Mass Gate Spike The spike occurs at

Página 238 - 5.1.4 Modifying an

Chapter 5 Optimizing Instrument Settings5-82 PerSeptive Biosystems5

Página 239 - Manual Control

6ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 6-16 Acquiring Mass SpectraThis chapter contains the following sections:6.1 Before You Be

Página 240 - 5-10 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-2 PerSeptive Biosystems66.1 Before You BeginThis section includes:• Overview of acquisition options• Guidelines

Página 241 - 5.1.5 Saving and Printing

Before You BeginVoyager™ Biospectrometry™ Workstation User’s Guide 6-36Automatic Control mode in Instrument Control Panel• Single sample• Automatic/ma

Página 242 - Printing

Chapter 6 Acquiring Mass Spectra6-4 PerSeptive Biosystems66.1.2 Guidelines for AcquiringThis section includes:• High voltage warmup for improved

Página 243 - Files to “Read-Only” Status

Before You BeginVoyager™ Biospectrometry™ Workstation User’s Guide 6-56When acquiring aspectrumWhen acquiring a spectrum, note the following:• For the

Página 244 - Parameter Descriptions

Chapter 6 Acquiring Mass Spectra6-6 PerSeptive Biosystems6• Observe the Spectrum window or Oscilloscope screen to determine if data is acceptable.

Página 245

Before You BeginVoyager™ Biospectrometry™ Workstation User’s Guide 6-766.1.3 Calibrating the Mass ScaleThis section describes:• Types of calibration•

Página 246 - 5-16 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-2 PerSeptive Biosystems11.1 Voyager™-DE and Voyager-DE PRO System OverviewThe

Página 247

Chapter 6 Acquiring Mass Spectra6-8 PerSeptive Biosystems6• Sequence Control calibration–Provides external, internal, and internal with automatic

Página 248 - 5-18 PerSeptive Biosystems

Before You BeginVoyager™ Biospectrometry™ Workstation User’s Guide 6-96CalibrationequationsThe general equation that the Voyager software uses for cal

Página 249

Chapter 6 Acquiring Mass Spectra6-10 PerSeptive Biosystems6GeneratedcalibrationIf you are performing an internal standard calibration, the softwar

Página 250 - 5-20 PerSeptive Biosystems

Acquiring in Manual Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-1166.2 Acquiring in Manual Mode from t

Página 251 - User’s Guide

Chapter 6 Acquiring Mass Spectra6-12 PerSeptive Biosystems6Figure 6-2 Instrument Control Panel Before AcquiringNOTE: If the Instrument Settings c

Página 252 - J. Am. Soc. Mass. Spectrom.

Acquiring in Manual Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-1363. Specify calibration information:•

Página 253 - Modifying the

Chapter 6 Acquiring Mass Spectra6-14 PerSeptive Biosystems6Hint: You can double-click on the control page to “undock” it, and click-drag it to any

Página 254 - Instrument Mode Tab Displayed

Acquiring in Manual Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-1563. Type a filename.4. Select Autoseq

Página 255

Chapter 6 Acquiring Mass Spectra6-16 PerSeptive Biosystems6DuringacquisitionDuring acquisition:• The Live/Current Spectrum trace in the Spectrum w

Página 256 - Voyager-DE system

Acquiring in Manual Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-176Figure 6-4 Checking for Signal Satu

Página 257

Voyager™-DE and Voyager-DE PRO System OverviewVoyager™ Biospectrometry™ Workstation User’s Guide 1-31Voyager-DE PROThe Voyager-DE™PRO Biospectrometry™

Página 258 - 5-28 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-18 PerSeptive Biosystems6Saving dataTo save the data when you have more than one trace displayed in the Spectrum

Página 259 - Advanced Tab Displayed

Acquiring in Manual Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-1966.2.2 Manually Accumulating Spectra

Página 260 - 5-30 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-20 PerSeptive Biosystems66.3 Obtaining Good Spectra in Delayed Extraction ModeThis section describes:• What is

Página 261

Obtaining Good Spectra in Delayed Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide 6-216Figure 6-6 Example of Poor Mass Spectrum for

Página 262 - 5-32 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-22 PerSeptive Biosystems6Figure 6-8 Example of Poor Mass Spectrum for AngiotensinFigure 6-9 Example of Good Ma

Página 263

Obtaining Good Spectra in Delayed Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide 6-2366.3.2 Laser IntensityOverviewLaser intensity

Página 264 - 5-34 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-24 PerSeptive Biosystems6Prompt fragmentsPrompt fragments appear in the spectrum at masses that correspond to th

Página 265

Making Accurate Mass MeasurementsVoyager™ Biospectrometry™ Workstation User’s Guide 6-2566.4 Making Accurate Mass MeasurementsOverviewAccurate mass m

Página 266 - 5-36 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-26 PerSeptive Biosystems6• Signal intensity of reference and analyte peaks is in the same range—Signal intensity

Página 267

Evaluating Data in the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-2766.5 Evaluating Data in the Instrument Control

Página 268 - 5-38 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-4 PerSeptive Biosystems1FeaturesFeatures of the Voyager-DE and Voyager-DE PRO

Página 269 - 5.3.1 Summary of Parameters

Chapter 6 Acquiring Mass Spectra6-28 PerSeptive Biosystems6NOTE: The parameters in this dialog box correspond to the Advanced Settings tab in the

Página 270 - 5-40 PerSeptive Biosystems

Evaluating Data in the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-2966.5.2 Calculating Mass ResolutionYou can calcu

Página 271

Chapter 6 Acquiring Mass Spectra6-30 PerSeptive Biosystems65. Type in up to four Mass/Charge values for which to calculate resolution.6. For each

Página 272 - Potential gradient

Evaluating Data in the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-316Figure 6-13 Resolution Calculator ResultsThe t

Página 273 - Adjusting

Chapter 6 Acquiring Mass Spectra6-32 PerSeptive Biosystems66.5.3 Calculating Signal-to-RMS Noise Ratio OverviewA signal-to-RMS noise ratio is typ

Página 274 - 5-44 PerSeptive Biosystems

Evaluating Data in the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-3364. For each Mass/Charge, enter the window for c

Página 275 - Grid Voltage%

Chapter 6 Acquiring Mass Spectra6-34 PerSeptive Biosystems66.6 Acquiring in Automatic Mode from the Instrument Control PanelThis section includes

Página 276 - Linear mode

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-356See Section 2.7, Aligning the Sam

Página 277 - Horizontal

Chapter 6 Acquiring Mass Spectra6-36 PerSeptive Biosystems6NOTE: Make sure the instrument settings yield acceptable results in Manual Control mode

Página 278 - 5-48 PerSeptive Biosystems

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-376Laser4. Select Use Automated Lase

Página 279 - Resolution and Data Quality

Voyager-DE STR System OverviewVoyager™ Biospectrometry™ Workstation User’s Guide 1-511.2 Voyager-DE STR System OverviewVoyager-DE STRThe Voyager™-DE

Página 280 - • Input Bandwidth

Chapter 6 Acquiring Mass Spectra6-38 PerSeptive Biosystems6Spectrumaccumulation andsaving8. Specify Spectrum Accumulation parameters:• Acquire X S

Página 281 - Suggested settings

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-396Sample positioning10. Select Use

Página 282 - 5-52 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-40 PerSeptive Biosystems6Setting spectrumacceptance andlaser adjustmentcriteriaIf you selected an accumulation c

Página 283 - Bandwidth setting

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-416Table 6-3 Spectrum Acceptance C

Página 284 - Settings Parameters

Chapter 6 Acquiring Mass Spectra6-42 PerSeptive Biosystems63. Click OK to close the Spectrum Acceptance Criteria dialog box.4. Click OK to close t

Página 285 - 5.4.1 Optimization Strategy

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-4366.6.3 Automatically Acquiring, E

Página 286 - Strategy

Chapter 6 Acquiring Mass Spectra6-44 PerSeptive Biosystems6DuringacquisitionDuring acquisition:• Laser is adjusted, sample position is adjusted, a

Página 287 - Laser intensity

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-456A search pattern file is an ASCII

Página 288 - General strategy

Chapter 6 Acquiring Mass Spectra6-46 PerSeptive Biosystems6Default searchpattern fileThe default search pattern file for a 100-position plate (DEF

Página 289 - Signal saturation

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-476Spiral searchpattern fileThe SPIR

Página 290 - 5-60 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-6 PerSeptive Biosystems1BiospectrometryBiospectrometry is the application of m

Página 291 - 5.4.3 Optimizing Resolution

Chapter 6 Acquiring Mass Spectra6-48 PerSeptive Biosystems6Hint: To determine coordinates, spot a sample plate with standard, load the plate, star

Página 292 - 5.4.3.1 Overview

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-4966.6.5 Process that Occurs During

Página 293

Chapter 6 Acquiring Mass Spectra6-50 PerSeptive Biosystems6In Prescan mode, the system does the following:1. The system sets the laser to the maxi

Página 294 - 5-64 PerSeptive Biosystems

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-5162. The system sets the laser to t

Página 295 - Delayed Extraction Mode

Chapter 6 Acquiring Mass Spectra6-52 PerSeptive Biosystems6Acquisition modeAcquisition mode starts after Prescan mode determines the laser setting

Página 296 - For more

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-536NOTE: If no minimum or maximum si

Página 297 - • Optimizing Delay Time

Chapter 6 Acquiring Mass Spectra6-54 PerSeptive Biosystems64. If Save All or Accumulate All is selected for accumulation, the system moves to the

Página 298 - 5-68 PerSeptive Biosystems

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-556If Acceptance Criteria are not me

Página 299 - Time and Grid Voltage% Values

Chapter 6 Acquiring Mass Spectra6-56 PerSeptive Biosystems66.6.6 Process that Occurs when Accumulating Spectra from Multiple Search Pattern Posit

Página 300 - 5-70 PerSeptive Biosystems

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-5766.6.6.1 Process that Occurs when

Página 301

MALDI-TOF MS Technology OverviewVoyager™ Biospectrometry™ Workstation User’s Guide 1-711.3 MALDI-TOF MS Technology OverviewMassspectrometryMass spect

Página 302 - 5-72 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-58 PerSeptive Biosystems66.6.6.2 Process that Occurs when Accumulating Passing SpectraWhen accumulating only th

Página 303

Acquiring in Automatic Mode from the Instrument Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-5969. Repeats step 2 through step 8

Página 304 - 5-74 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-60 PerSeptive Biosystems66.7 Acquiring Spectra from the Sequence Control PanelOverviewThe Voyager Sequence Cont

Página 305

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-616Figure 6-19 Sequence Control Panel

Página 306 - 5.4.4.1 Overview

Chapter 6 Acquiring Mass Spectra6-62 PerSeptive Biosystems66.7.1 Understanding Settings, Macros, and CalibrationThis section includes:• File type

Página 307

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-636How file typesand calibrationspecified affect

Página 308 - 5-78 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-64 PerSeptive Biosystems6Figure 6-20 Settings, Macros, and Calibration in Sequence ControlNoCalibrationDataExte

Página 309

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-6566.7.2 Before Creating a SequenceThis section

Página 310 - 5-80 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-66 PerSeptive Biosystems6Creating macrosYou can specify macros that execute before and after calibration. You ca

Página 311

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-676Creating settings(.SET) filesCreate settings

Página 312 - 5-82 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-8 PerSeptive Biosystems1Matrix-AssistedLaser DesorptionIonization (MALDI)In Ma

Página 313 - Mass Spectra

Chapter 6 Acquiring Mass Spectra6-68 PerSeptive Biosystems66.7.3.1 Setting General Sequence ParametersSet parameters as needed:1. From the View m

Página 314 - 6.1 Before You Begin

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-696NOTE: Run logs are overwritten each time you

Página 315 - Workstation User’s Guide 6-3

Chapter 6 Acquiring Mass Spectra6-70 PerSeptive Biosystems6Figure 6-22 Sequence Run ListHint: You can show and hide columns by selecting a column

Página 316 - 6-4 PerSeptive Biosystems

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-716Instrument Settings File(required entry) .BIC

Página 317 - Workstation User’s Guide 6-5

Chapter 6 Acquiring Mass Spectra6-72 PerSeptive Biosystems6Internal/External CalibrationCalibration type to use for this row. Click the down arrow

Página 318 - 6-6 PerSeptive Biosystems

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-736Calibration File(required entry for External,

Página 319 - Types of

Chapter 6 Acquiring Mass Spectra6-74 PerSeptive Biosystems65. When all necessary rows are filled in, save the sequence by selecting Save Sequence

Página 320 - When to calibrate

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-756Modifying andcustomizing therun listYou can m

Página 321 - Workstation User’s Guide 6-9

Chapter 6 Acquiring Mass Spectra6-76 PerSeptive Biosystems6ImportingTo import a .TXT or .XLS file:1. Select Import from the File menu.2. Select a

Página 322 - Software User’s Guide

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-7766.7.4 Preparing to Run a SequenceThis sectio

Página 323 - 6.2.1 Manually Acquiring

MALDI-TOF MS Technology OverviewVoyager™ Biospectrometry™ Workstation User’s Guide 1-91For acquiring TOF spectra, time measurement starts:• Delayed Ex

Página 324 - 6-12 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-78 PerSeptive Biosystems6Before acquiringBefore acquiring a sequence:1. Load a sequence by doing one of the foll

Página 325

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-796Pausing andresuming asequenceTo pause a seque

Página 326 - 6-14 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-80 PerSeptive Biosystems66.7.6 Checking Sequence StatusYou can check sequence status in three places:• Acquisit

Página 327

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-816Sequence status parameters include:Checking s

Página 328 - 6-16 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-82 PerSeptive Biosystems66.7.7 Automatic Calibration During a Sequence RunThis section includes:• Calibration o

Página 329

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-8366.7.7.2 Calibration Standard RequirementsMas

Página 330 - 6-18 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-84 PerSeptive Biosystems6Sample andstandard inseparate samplepositionsIf you are acquiring samples and standards

Página 331

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-856The number and placement of standards needed

Página 332 - 6.3 Obtaining Good Spectra

Chapter 6 Acquiring Mass Spectra6-86 PerSeptive Biosystems6To acquire sample and standard, create two search pattern files to analyze the sample a

Página 333 - Lower laser power producing:

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-876NOTE: Use plates without laser-etched sample

Página 334

Table of Contentsiv PerSeptive BiosystemsChapter 2 Installing the Voyager™ Biospectrometry™ Workstations2.1 Installing the System ...

Página 335 - 6.3.2 Laser Intensity

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-10 PerSeptive Biosystems1Drift time is proportional to the square root of the

Página 336 - 6.3.3 Parameters Affecting

Chapter 6 Acquiring Mass Spectra6-88 PerSeptive Biosystems6If standard signal suppresses unknown signal, you have two options for analysis:• Perfo

Página 337 - Mass Measurements

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-8962. Load the sample plate, select the position

Página 338 - 6-26 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-90 PerSeptive Biosystems66.7.8 Customizing the Sequence DisplayUsing WorkbookmodeWorkbook mode displays the run

Página 339 - 6.5.1 Detecting

Acquiring Spectra from the Sequence Control PanelVoyager™ Biospectrometry™ Workstation User’s Guide 6-9165. To set preferences, select Preferences fro

Página 340 - 6-28 PerSeptive Biosystems

Chapter 6 Acquiring Mass Spectra6-92 PerSeptive Biosystems6

Página 341 - Height for most applications

7ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 7-17 PSD AnalysisThis chapter contains the following sections:7.1 Overview of PSD Analysi

Página 342 - 6-30 PerSeptive Biosystems

Chapter 7 PSD Analysis7-2 PerSeptive Biosystems77.1 Overview of PSD AnalysisThis section includes:• Post-source decay analysis• Segments and comp

Página 343 - ____ ____

Overview of PSD AnalysisVoyager™ Biospectrometry™ Workstation User’s Guide 7-37PSD fragmentionsAt higher laser intensities, some molecular ions decomp

Página 344 - 6-32 PerSeptive Biosystems

Chapter 7 PSD Analysis7-4 PerSeptive Biosystems7Figure 7-1 Molecular and Fragment Ion Flight in the ReflectorThe AH+ fragment has 70 percent of t

Página 345 - 5. Click OK

Overview of PSD AnalysisVoyager™ Biospectrometry™ Workstation User’s Guide 7-57Figure 7-2 Effect of Changing Mirror RatioA Mirror Ratio setting of 1.

Página 346 - Automatic Control Mode

Voyager-DE™ (Delayed Extraction™) TechnologyVoyager™ Biospectrometry™ Workstation User’s Guide 1-1111.4 Voyager-DE™ (Delayed Extraction™) TechnologyI

Página 347

Chapter 7 PSD Analysis7-6 PerSeptive Biosystems77.1.2 Segments and Composite SpectraOverviewTo obtain the most information about an ion, collect

Página 348 - Control mode

Overview of PSD AnalysisVoyager™ Biospectrometry™ Workstation User’s Guide 7-777.1.3 PSD Data FilesPSD data (.DAT) files include:• Precursor ion mass

Página 349

Chapter 7 PSD Analysis7-8 PerSeptive Biosystems77.1.4 Mass Calculation for Fragment IonsPSD calibrationequationPSD calibration specifies the mass

Página 350 - Spectrum

Overview of PSD AnalysisVoyager™ Biospectrometry™ Workstation User’s Guide 7-977.1.5 Optimizing the Precursor Ion SelectorThe Precursor Ion Selector

Página 351

Chapter 7 PSD Analysis7-10 PerSeptive Biosystems7Figure 7-4 Timed Ion Selector Tab in Hardware Configuration Dialog BoxSet the Deflector Gate Wid

Página 352 - 6-40 PerSeptive Biosystems

Enhancing Fragmentation with CIDVoyager™ Biospectrometry™ Workstation User’s Guide 7-1177.2 Enhancing Fragmentation with CIDOverviewCollision-induced

Página 353 - Continued

Chapter 7 PSD Analysis7-12 PerSeptive Biosystems7BenefitsThe benefits provided by CID include: • Fragmentation of ions that does not occur under n

Página 354 - Saving the

Enhancing Fragmentation with CIDVoyager™ Biospectrometry™ Workstation User’s Guide 7-137Figure 7-7 Mid Masses—Impact of CID (Glu-1-Fibrinopeptide)In

Página 355

Chapter 7 PSD Analysis7-14 PerSeptive Biosystems7Purging collisiongas linesBefore turning on the collision gas, purge the lines to prevent disrupt

Página 356 - 6.6.4 Search Pattern Files

Enhancing Fragmentation with CIDVoyager™ Biospectrometry™ Workstation User’s Guide 7-157Turning oncollision gasTo turn on the collision gas after purg

Página 357 - 0,0 1270,0-1270,0

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-12 PerSeptive Biosystems1DelayedExtractionWith Voyager-DE™ (Delayed Extraction

Página 358 - Default search

Chapter 7 PSD Analysis7-16 PerSeptive Biosystems7Turning offcollision gasTo turn off the collision gas:1. Turn the top 3-way valve on the CID box

Página 359

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-1777.3 Acquiring PSD Data with

Página 360 - 6-48 PerSeptive Biosystems

Chapter 7 PSD Analysis7-18 PerSeptive Biosystems7Differences fromregular analysisWhen operating in PSD mode, note the following differences from a

Página 361 - Acquisition in Automatic Mode

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-1977.3.1 Determining the Precur

Página 362 - 6-50 PerSeptive Biosystems

Chapter 7 PSD Analysis7-20 PerSeptive Biosystems7Generating anexternalcalibration for theprecursor ionTo obtain maximum mass accuracy for the prec

Página 363

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-2177.3.2 Determining the Number

Página 364 - 6-52 PerSeptive Biosystems

Chapter 7 PSD Analysis7-22 PerSeptive Biosystems7Hint: You can use the Windows calculator to determine natural log values. If the ln function is n

Página 365 - Control dialog box is used

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-237Size of segmentsYou can acqui

Página 366 - 6-54 PerSeptive Biosystems

Chapter 7 PSD Analysis7-24 PerSeptive Biosystems77.3.3 Setting PSD Acquisition ParametersSetting PSD Acquisition parameters includes:• Displaying

Página 367

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-2572. If the PSD Acquisition Set

Página 368 - 6.6.6 Process that Occurs

Voyager-DE™ (Delayed Extraction™) TechnologyVoyager™ Biospectrometry™ Workstation User’s Guide 1-131In Continuous Extraction mode:• Accelerating volta

Página 369 - Accumulating All Spectra

Chapter 7 PSD Analysis7-26 PerSeptive Biosystems74. In the Calibration section of the Instrument Settings control page, select the matrix you are

Página 370 - Accumulating Passing Spectra

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-277NOTE: The calibration you spe

Página 371

Chapter 7 PSD Analysis7-28 PerSeptive Biosystems77.3.4 Filling in the Segment List This section describes:• Filling in the list• Typing or select

Página 372 - 6.7 Acquiring Spectra from

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-297Rows are added with the follo

Página 373

Chapter 7 PSD Analysis7-30 PerSeptive Biosystems7Typing orselecting newvaluesTo type or select new values in the segment list:1. Type in new value

Página 374 - Macros, and Calibration

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-317NOTE: Do not select Guide Wir

Página 375 - How file types

Chapter 7 PSD Analysis7-32 PerSeptive Biosystems77.3.5 Acquiring and Saving PSD SegmentsThis section includes:•Overview• Acquiring PSD segments•

Página 376 - 6-64 PerSeptive Biosystems

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-337NOTE: If you stop an experime

Página 377

Chapter 7 PSD Analysis7-34 PerSeptive Biosystems7DuringacquisitionThe following occurs:• Mass range for the segment is set to:• Acquisition starts

Página 378 - 6-66 PerSeptive Biosystems

Acquiring PSD Data with Standard Instrument Settings (.BIC) FilesVoyager™ Biospectrometry™ Workstation User’s Guide 7-357Accumulating orsaving thesegm

Página 379 - 6.7.3 Creating a Sequence

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-14 PerSeptive Biosystems1Benefits ofDelayedExtractionDelayed Extraction of ion

Página 380 - 6-68 PerSeptive Biosystems

Chapter 7 PSD Analysis7-36 PerSeptive Biosystems77.4 Exploring PSD ModeTo be successful in PSD analysis, you need to understand how ions behave i

Página 381 - 6.7.3.2 Creating a Run List

Exploring PSD ModeVoyager™ Biospectrometry™ Workstation User’s Guide 7-3777.4.1 Observing the Effects of Laser IntensityAdjusting laser intensity aff

Página 382 - View menu

Chapter 7 PSD Analysis7-38 PerSeptive Biosystems75. Note the behavior of the signal intensity for the first few spectra and subsequent continued s

Página 383

Exploring PSD ModeVoyager™ Biospectrometry™ Workstation User’s Guide 7-397Hint: As you initially experiment, adjust the laser in large steps, for exam

Página 384 - 6-72 PerSeptive Biosystems

Chapter 7 PSD Analysis7-40 PerSeptive Biosystems77.4.2 Observing the Effects of Precursor Ion SelectorTurning on the Precursor Ion Selector elimi

Página 385

Exploring PSD ModeVoyager™ Biospectrometry™ Workstation User’s Guide 7-417Figure 7-12 compares the spectrum above with the spectrum acquired in the pr

Página 386 - 6-74 PerSeptive Biosystems

Chapter 7 PSD Analysis7-42 PerSeptive Biosystems7Observing promptand PSDfragmentsWhen operating under conditions that yield high resolution for an

Página 387

Exploring PSD ModeVoyager™ Biospectrometry™ Workstation User’s Guide 7-437Figure 7-13 Flight Path of Prompt and PSD FragmentsPath of prompt fragment

Página 388 - 6-76 PerSeptive Biosystems

Chapter 7 PSD Analysis7-44 PerSeptive Biosystems77.4.3 Observing the Effects of Grid Voltage% Adjusting Grid Voltage% affects resolution.In this

Página 389

Exploring PSD ModeVoyager™ Biospectrometry™ Workstation User’s Guide 7-457By fine-tuning the Grid Voltage%, you can alter the point of initial time fo

Página 390 - 6.7.5 Running a Sequence

Voyager-DE™ (Delayed Extraction™) TechnologyVoyager™ Biospectrometry™ Workstation User’s Guide 1-151Velocity focusingDelayed Extraction technology fac

Página 391

Chapter 7 PSD Analysis7-46 PerSeptive Biosystems7AcquiringNOTE: In this exercise, you will observe the impact of the Grid Voltage% on the precurso

Página 392 - 6-80 PerSeptive Biosystems

Exploring PSD ModeVoyager™ Biospectrometry™ Workstation User’s Guide 7-4776. Check the Grid Voltage% setting in the standard .BIC file provided on you

Página 393 - Control Panel

Chapter 7 PSD Analysis7-48 PerSeptive Biosystems77.5 Viewing PSD DataFor information on viewing PSD data, see the Data Explorer Software User’s G

Página 394 - During a Sequence Run

8ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 8-138 Maintenance andTroubleshootingThis chapter contains the following sections:8.1 Main

Página 395

Chapter 8 Maintenance and Troubleshooting8-2 PerSeptive Biosystems88.1 MaintenanceThis section describes:• Maintenance schedule• Hardware mainten

Página 396 - Optimum Mass Accuracy

MaintenanceVoyager™ Biospectrometry™ Workstation User’s Guide 8-388.1.2 Hardware MaintenanceWARNINGELECTRICAL SHOCK HAZARD. Severe electrical shock c

Página 397

Chapter 8 Maintenance and Troubleshooting8-4 PerSeptive Biosystems8• Inspect flap valve 1, flap valve 2, linear actuator o-rings• Clean optics and

Página 398 - 6-86 PerSeptive Biosystems

MaintenanceVoyager™ Biospectrometry™ Workstation User’s Guide 8-58Figure 8-1 Changing FusesWARNINGFIRE HAZARD. Using a fuse of the wrong type or rati

Página 399 - Calibration Considerations

Chapter 8 Maintenance and Troubleshooting8-6 PerSeptive Biosystems86. Insert two fuses of the proper rating.7. Insert the voltage selector/fuse ho

Página 400 - 6-88 PerSeptive Biosystems

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-788.2 TroubleshootingThis section includes:• Spectrum troubleshooting• Software t

Página 401 - Using Separate Spots

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-16 PerSeptive Biosystems1Figure 1-10 Velocity Focusing of Ions in Linear Mode

Página 402 - Using Workbook

Chapter 8 Maintenance and Troubleshooting8-8 PerSeptive Biosystems8Flat signal on oscilloscope or in spectrum window for sample region (matrix pea

Página 403 - Arranging

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-98Flat signal on oscilloscope or in spectrum window for sample and matrix regionVe

Página 404 - 6-92 PerSeptive Biosystems

Chapter 8 Maintenance and Troubleshooting8-10 PerSeptive Biosystems8Poor resolution/ sensitivity in Delayed Extraction modeDelay Time and Grid Vol

Página 405 - 7 PSD Analysis

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-118Round tops on peaks Saturated (flat top) and unsaturated (sharp top) scans aver

Página 406 - 7.1 Overview of PSD Analysis

Chapter 8 Maintenance and Troubleshooting8-12 PerSeptive Biosystems8Poor mass accuracy in Delayed Extraction mode(external calibration only)High v

Página 407 - Workstation User’s Guide 7-3

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-138Poor signal-to-noise ratio or sensitivity(continued on next page)Laser intensit

Página 408 - Reflector

Chapter 8 Maintenance and Troubleshooting8-14 PerSeptive Biosystems8Poor signal-to-noise ratio or sensitivity(continued)Guide Wire Voltage% too hi

Página 409 - J. Am. Soc. Mass Spectrom.

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-158Dimer in spectrum Laser intensity too highAdjust laser by using the slider cont

Página 410 - 7-6 PerSeptive Biosystems

Chapter 8 Maintenance and Troubleshooting8-16 PerSeptive Biosystems8Peaks not symmetrical Laser intensity too high Decrease laser intensity by usi

Página 411 - 7.1.3 PSD Data Files

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-178On Voyager-DE PRO and Voyager-DE STR systems, cannot see high mass ions in Refl

Página 412 - Default PSD

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-1711.5 Parts of the Voyager-DE andVoyager-DE P

Página 413 - Workstation User’s Guide 7-9

Chapter 8 Maintenance and Troubleshooting8-18 PerSeptive Biosystems8Large tail on the high mass side of peakUnresolved salt or buffer adducts due

Página 414 - Configuration Dialog Box

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-1988.2.2 Software TroubleshootingThis section includes:• Instrument Control Panel

Página 415 - Fragmentation with CID

Chapter 8 Maintenance and Troubleshooting8-20 PerSeptive Biosystems8Calibration (mass) shifted up or down by 10 Da Uneven matrix layer causing hot

Página 416 - Benefits

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-218Checking theWindows NTEvent LogThe Windows NT Event Log is a running list of ev

Página 417 - Meth. Enzymol

Chapter 8 Maintenance and Troubleshooting8-22 PerSeptive Biosystems88.2.3 Hardware TroubleshootingThis section includes:• Mass spectrometer troub

Página 418 - 7-14 PerSeptive Biosystems

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-238Cracking sound in mass spectrometerArcing caused by dirty sample plateUse clean

Página 419

Chapter 8 Maintenance and Troubleshooting8-24 PerSeptive Biosystems8Sample holder empty when you click EjectSample plate too far way from grabber

Página 420 - 7-16 PerSeptive Biosystems

TroubleshootingVoyager™ Biospectrometry™ Workstation User’s Guide 8-258E09 error message displayed on gauge controller for BA1 or BA2BA1 or BA2 gauges

Página 421 - Settings (.BIC) Files

Chapter 8 Maintenance and Troubleshooting8-26 PerSeptive Biosystems8

Página 422 - Differences from

AAppendixVoyager™ Biospectrometry™ Workstation User’s Guide A-1A SpecificationsThis appendix contains the following sections:A.1 Voyager-DE Specific

Página 423 - Generating the

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-18 PerSeptive Biosystems1Voyager-DE PROThe Voyager-DE PRO Biospectrometry Work

Página 424 - 7-20 PerSeptive Biosystems

Appendix A SpecificationsA-2 PerSeptive BiosystemsAA.1 Voyager-DE SpecificationsThis section includes the following specifications for the Voyage

Página 425 - 7.3.2 Determining the

Voyager-DE SpecificationsVoyager™ Biospectrometry™ Workstation User’s Guide A-3AIon source voltages Tunable:• Accelerating Voltage—Up to 25,000 V• Gri

Página 426 - 7-22 PerSeptive Biosystems

Appendix A SpecificationsA-4 PerSeptive BiosystemsAA.2 Voyager-DE PRO SpecificationsThis section includes the following specifications for the Vo

Página 427

Voyager-DE PRO SpecificationsVoyager™ Biospectrometry™ Workstation User’s Guide A-5ATable A-5 Voyager-DE PRO Mass Spectrometer SpecificationsConditi

Página 428 - 7-24 PerSeptive Biosystems

Appendix A SpecificationsA-6 PerSeptive BiosystemsATable A-6 Voyager-DE PRO Miscellaneous SpecificationsCondition SpecificationOperating tempera

Página 429

Voyager-DE STR SpecificationsVoyager™ Biospectrometry™ Workstation User’s Guide A-7AA.3 Voyager-DE STR SpecificationsThis section includes the follow

Página 430 - 7-26 PerSeptive Biosystems

Appendix A SpecificationsA-8 PerSeptive BiosystemsATable A-8 Voyager-DE STR Mass Spectrometer SpecificationsCondition SpecificationReflector Sin

Página 431

Voyager-DE STR SpecificationsVoyager™ Biospectrometry™ Workstation User’s Guide A-9ATable A-9 Voyager-DE STR Miscellaneous SpecificationsCondition S

Página 432 - 7-28 PerSeptive Biosystems

Appendix A SpecificationsA-10 PerSeptive BiosystemsA

Página 433

BAppendixVoyager™ Biospectrometry™ Workstation User’s Guide B-1B Warranty/ServiceInformationThis appendix contains the following sections:B.1 Limite

Página 434 - 7-30 PerSeptive Biosystems

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-191Parts of the Voyager-DE and Voyager-DE PRO s

Página 435

Appendix B Warranty/Service InformationB-2 PerSeptive BiosystemsBB.1 Limited Product WarrantyLimited warrantyPerSeptive Biosystems, Inc. (“PerSep

Página 436 - 7-32 PerSeptive Biosystems

Voyager™ Biospectrometry™ Workstation User’s Guide B-3BPerSeptive warrants that for a period of ninety (90) days the software designated for use with

Página 437

Appendix B Warranty/Service InformationB-4 PerSeptive BiosystemsBWarrantylimitationsTHE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE RE

Página 438 - Changing

Voyager™ Biospectrometry™ Workstation User’s Guide B-5BB.2 Damages, Claims, ReturnsDamagesPlease examine any shipments promptly after receipt to chec

Página 439

Appendix B Warranty/Service InformationB-6 PerSeptive BiosystemsBB.3 Spare PartsStandards andmatricesThe following compounds are available from t

Página 440 - 7.4 Exploring PSD Mode

Voyager™ Biospectrometry™ Workstation User’s Guide B-7BSample platesThe following sample plates are available from PerSeptive Biosystems:Description P

Página 441 - Effects of Laser Intensity

Appendix B Warranty/Service InformationB-8 PerSeptive BiosystemsBMembrane, Gels V700698Hydrophobic plastic surface, flat, 400-position V700699Desc

Página 442 - 7-38 PerSeptive Biosystems

CAppendixVoyager™ Biospectrometry™ Workstation User’s Guide C-1C MatricesThis appendix lists commonly used matrices. For each matrix, it also lists:•

Página 443 - Exploring PSD Mode

Appendix C MatricesC-2 PerSeptive BiosystemsCFigure C-1 Sinapinic Acid Matrix SpectrumFigure C-2 α-cyano-4-hydroxycinnamic acid (CHCA) Matrix Sp

Página 444 - 7.4.2 Observing the Effects

Voyager™ Biospectrometry™ Workstation User’s Guide C-3CFigure C-3 2,5-dihydroxybenzoic acid (2,5-DHB) Matrix SpectrumFigure C-4 Mixture of 2,5-dihyd

Página 445

Table of ContentsVoyager Biospectrometry Workstation User’s Guide vChapter 3 Preparing Samples3.1 Preparing Samples ...

Página 446 - 7-42 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-20 PerSeptive Biosystems11.5.2 Mass SpectrometerVoyager-DEThe parts of the Vo

Página 447

Appendix C MatricesC-4 PerSeptive BiosystemsCFigure C-5 2-(4-hydroxy-phenylazo)-benzoic acid (HABA) Matrix SpectrumFigure C-6 3-hydroxypicolinic

Página 448 - Figure 7-14 Ion Focusing

Voyager™ Biospectrometry™ Workstation User’s Guide C-5CFigure C-7 Dithranol Matrix SpectrumFigure C-8 2,4,6 trihydroxyacetophenone (THAP) Spectrum

Página 449

Appendix C MatricesC-6 PerSeptive BiosystemsCFigure C-9 trans-3-indoleacrylic acid (IAA) Matrix SpectrumMatrix Applications/ColorMatrix Solution

Página 450 - 7-46 PerSeptive Biosystems

Voyager™ Biospectrometry™ Workstation User’s Guide C-7CAlpha-cyano-4-hydroxycinnamic acid (αCHCA)(see mass spectrum on page C-2)MW 189.04 DaApplicatio

Página 451 - 7.4.4 Summary

Appendix C MatricesC-8 PerSeptive BiosystemsC2,5-dihydroxybenzoic acid (2,5-DHB)(see mass spectrum on page C-3)MW 154.03 DaApplications:Small mole

Página 452 - 7.5 Viewing PSD Data

Voyager™ Biospectrometry™ Workstation User’s Guide C-9Cmixture of 2,5-dihydroxybenzoic acid and 5-methoxysalicylic acid (DHBs)(see mass spectrum on pa

Página 453 - Troubleshooting

Appendix C MatricesC-10 PerSeptive BiosystemsC2-(4-hydroxy-phenylazo)-benzoic acid (HABA)(see mass spectrum on page C-4)MW 242.07 DaApplications:•

Página 454 - 8.1.1 Maintenance Schedule

Voyager™ Biospectrometry™ Workstation User’s Guide C-11CDithranol(see mass spectrum on page C-5)MW 226.06 DaApplications:Nonpolar synthetic polymersCo

Página 455 - 8.1.2 Hardware Maintenance

Appendix C MatricesC-12 PerSeptive BiosystemsCtrans-3-indoleacrylic acid (IAA) (see mass spectrum on page C-6)MW 187.2Applications:Non-polar polym

Página 456 - Changing fuses

DAppendixVoyager™ Biospectrometry™ Workstation User’s Guide D-1D Log SheetsThis appendix includes log sheets you can copy and use to log samples befo

Página 457 - Figure 8-1 Changing Fuses

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-211Voyager-DE PROThe parts of the Voyager-DE PR

Página 458 - (5 x 20 mm)

AVoyager™ Biospectrometry™ Sample LogPlate #: Date:1234567891012345678910

Página 459 - 8.2 Troubleshooting

A.Voyager™ Biospectrometry™ Sample Log Plate: Date: Page of Path and File NameSamp # Matrix Sample Linear Reflector1.2.3.4.5.

Página 460 - 8-8 PerSeptive Biosystems

A26.27.28.29.30.31.32.33.34.35.36.37.38.39.40.41.42.43.44.45.46.47.48.49.50.Voyager™ Biospectrometry™ Sample Log Plate: Date: Page of

Página 461 - Workstation User’s Guide 8-9

A51.52.53.54.55.56.57.58.59.60.61.62.63.64.65.66.67.68.69.70.71.72.73.74.75.Voyager™ Biospectrometry™ Sample Log Plate: Date: Page of

Página 462 - NOTE: Input Bandwidth is not

A76.77.78.79.80.81.82.83.84.85.86.87.88.89.90.91.92.93.94.95.96.97.98.99.100.Voyager™ Biospectrometry™ Sample Log Plate: Date: Page of

Página 463 - NOTE: Data acquired using

AVoyager™ Biospectrometry™ Sample Log Plate: Date: Page of Path and File NameSamp # Matrix Sample Linear Reflector

Página 464 - 8-12 PerSeptive Biosystems

Appendix D Log SheetsD-8 PerSeptive BiosystemsD

Página 465 - (continued on next page)

EAppendixVoyager™ Biospectrometry™ Workstation User’s Guide E-1E Grid Voltage%and DelayTime SettingsThe following figures illustrate the relationship

Página 466 - (continued)

Appendix E Grid Voltage% and Delay Time SettingsE-2 PerSeptive BiosystemsE88 90 92 94 960200400600G rid Voltage (% )Pulse Delay (nsec)m/z=1000m/z=

Página 467

FAppendixVoyager™ Biospectrometry™ Workstation User’s Guide F-1F ReferenceStandardInformationThis appendix contains the following sections:F.1 Calib

Página 468 - NOTE: Increasing the laser

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-22 PerSeptive Biosystems1Parts of the massspectrometerThe Voyager-DE and Voyag

Página 469

Appendix F Reference Standard InformationF-2 PerSeptive BiosystemsFF.1 Calibration CompoundsThe table below includes masses for common calibratio

Página 470 - 8-18 PerSeptive Biosystems

Calibration CompoundsVoyager™ Biospectrometry™ Workstation User’s Guide F-3FInsulin B chain, oxidized3493.6435 3495.95 +1 3494.6513 3496.96Adrenocorti

Página 471

Appendix F Reference Standard InformationF-4 PerSeptive BiosystemsFF.2 Conversion of Mass to Time for Typical StandardsThe table below includes m

Página 472 - 8-20 PerSeptive Biosystems

Conversion of Mass to Time for Typical StandardsVoyager™ Biospectrometry™ Workstation User’s Guide F-5FBovine Insulin B, Oxidized1 3496.9 3.50E-05 5.5

Página 473 - Event Log

Appendix F Reference Standard InformationF-6 PerSeptive BiosystemsFF.3 Theoretical Cleavages for AngiotensinMonoisotopic masses (Da) for the theo

Página 474 - 8-22 PerSeptive Biosystems

Observed PSD Fragments in AngiotensinVoyager™ Biospectrometry™ Workstation User’s Guide F-7FF.4 Observed PSD Fragments in AngiotensinThe masses liste

Página 475

Appendix F Reference Standard InformationF-8 PerSeptive BiosystemsF272.136 272.3 b2 1000.537 1001.2 a8285.135 285.3 FH 1046.542 1047.2 b8+H2O326.1

Página 476 - 8-24 PerSeptive Biosystems

GAppendixVoyager™ Biospectrometry™ Workstation User’s Guide G-1G Maintenance LogThe following page includes a log sheet listing maintenance procedur

Página 477

Maintenance Log for Voyager System Serial Number ____________Record the date and your initials when you perform maintenance procedures.Weekly Jan

Página 478 - 8-26 PerSeptive Biosystems

HAppendixVoyager™ Biospectrometry™ Workstation User’s Guide H-1H ContinuousExtraction ModeThis appendix contains the following sections:H.1 Optimizi

Página 479 - Appendix

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-231• Linear detector—A device that detects ions

Página 480 - Appendix A Specifications

Appendix H Continuous Extraction ModeH-2 PerSeptive BiosystemsHH.1 Optimizing a Continuous Extraction Standard Instrument (.BIC) SettingBefore yo

Página 481 - Workstation User’s Guide A-3

Optimizing a Continuous Extraction Standard Instrument (.BIC) SettingVoyager™ Biospectrometry™ Workstation User’s Guide H-3HList of standardinstrument

Página 482 - Specifications

Appendix H Continuous Extraction ModeH-4 PerSeptive BiosystemsHTable H-2 Continuous Extraction Reflector Mode Standard Instrument Settings (.BIC

Página 483 - Workstation User’s Guide A-5

Optimizing a Continuous Extraction Standard Instrument (.BIC) SettingVoyager™ Biospectrometry™ Workstation User’s Guide H-5HOptimizationstrategyWhen o

Página 484

Appendix H Continuous Extraction ModeH-6 PerSeptive BiosystemsH6. Set the Grid Voltage% appropriate for the matrix and mass:7. Save the instrument

Página 485 - Workstation User’s Guide A-7

Optimizing a Continuous Extraction Standard Instrument (.BIC) SettingVoyager™ Biospectrometry™ Workstation User’s Guide H-7HTable H-5 Guide Wire Vol

Página 486

Appendix H Continuous Extraction ModeH-8 PerSeptive BiosystemsHH.2 Obtaining Good Spectra in Continuous Extraction ModeThis section describes:• S

Página 487 - Workstation User’s Guide A-9

Obtaining Good Spectra in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-9HFigure H-1 Example of Poor Mass Spectrum f

Página 488

Appendix H Continuous Extraction ModeH-10 PerSeptive BiosystemsHFigure H-3 Example of Poor Mass Spectrum for AngiotensinFigure H-4 Example of Go

Página 489

Obtaining Good Spectra in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-11HParametersaffectingresolution andsignal-to

Página 490 - B.1 Limited Product Warranty

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-24 PerSeptive Biosystems1The single-stage design provides high mass resolution

Página 491 - Workstation User’s Guide B-3

Appendix H Continuous Extraction ModeH-12 PerSeptive BiosystemsHH.2.2 Determining Laser Threshold This section describes:•Overview• Adjusting the

Página 492 - B-4 PerSeptive Biosystems

Obtaining Good Spectra in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-13HDeterminingthresholdDetermining laser thre

Página 493 - B.2 Damages, Claims, Returns

Appendix H Continuous Extraction ModeH-14 PerSeptive BiosystemsHFigure H-5 Saturated Spectrum3. Check the spectrum for the peak of interest. Zoom

Página 494 - B.3 Spare Parts

Obtaining Good Spectra in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-15HFine-tuningthresholdWhen you determine the

Página 495 - Sample plates

Appendix H Continuous Extraction ModeH-16 PerSeptive BiosystemsHH.2.2.3 Verifying Threshold SettingMove to a new area of the sample well that con

Página 496 - Description Part Number

Obtaining Good Spectra in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-17HH.2.3 Checking ResolutionAfter you determ

Página 497

Appendix H Continuous Extraction ModeH-18 PerSeptive BiosystemsHH.2.4 Fine-Tuning the Laser SettingWhen you find the laser threshold, whether you

Página 498 - Figure C-2

Troubleshooting in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-19HH.3 Troubleshooting in ContinuousExtraction Mode

Página 499 - (DHBs) Matrix Spectrum

Appendix H Continuous Extraction ModeH-20 PerSeptive BiosystemsHRefer to the table below if you are having spectrum trouble:Signal is flat Laser s

Página 500 - C-4 PerSeptive Biosystems

Troubleshooting in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-21HPoor resolution in Continuous Extraction mode(con

Página 501 - Workstation User’s Guide C-5

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-2511.5.3.1 Voyager-DE Vacuum SystemVacuumchamb

Página 502 - if sample

Appendix H Continuous Extraction ModeH-22 PerSeptive BiosystemsHPoor resolution in Continuous Extraction mode(continued)Beam guide wire malfunctio

Página 503 - Workstation User’s Guide C-7

Troubleshooting in Continuous Extraction ModeVoyager™ Biospectrometry™ Workstation User’s Guide H-23HPoor mass accuracy in Continuous Extraction mode(

Página 504 - PB100253

Appendix H Continuous Extraction ModeH-24 PerSeptive BiosystemsHPoor mass accuracy in Continuous Extraction mode(continued)Crystals did not form h

Página 505 - Workstation User’s Guide C-9

IAppendixVoyager™ Biospectrometry™ Workstation User’s Guide I-1I Using the Oscilloscope and Control StickThis appendix contains the following section

Página 506 - C-10 PerSeptive Biosystems

Appendix I Using the Oscilloscope and Control StickI-2 PerSeptive BiosystemsIOscilloscopeoverviewAn external oscilloscope (instead of the internal

Página 507

Guidelines for AcquiringVoyager™ Biospectrometry™ Workstation User’s Guide I-3II.1 Guidelines for AcquiringConsider the following as you acquire a sp

Página 508 - C-12 PerSeptive Biosystems

Appendix I Using the Oscilloscope and Control StickI-4 PerSeptive BiosystemsII.2 ScalingInitial scalingInitial scaling of the oscilloscope is det

Página 509

ScalingVoyager™ Biospectrometry™ Workstation User’s Guide I-5IUse these knobs to adjust the signal on the oscilloscope:• Vertical position—Moves the s

Página 510 - Sample Log

Appendix I Using the Oscilloscope and Control StickI-6 PerSeptive BiosystemsIThe Channel indicator (Ch1 in Figure I-3) displays Ch1 in Linear mode

Página 511

Using the Control StickVoyager™ Biospectrometry™ Workstation User’s Guide I-7II.3 Using the Control StickStartingacquisitionAfter you load samples on

Página 512

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-26 PerSeptive Biosystems1Vacuum pumpsTwo vacuum pumps create the vacuum enviro

Página 513

Appendix I Using the Oscilloscope and Control StickI-8 PerSeptive BiosystemsICAUTIONCheck to see if acquisition has already stopped automatically

Página 514

Voyager™ Biospectrometry™ Workstation User’s Guide Bibliography-1BBLIIOGRAPHYBibliographyGeneral Mass SpectrometryBeavis, R.C. and B.T. Chait, Chem. P

Página 515

Bibliography-2 PerSeptive BiosystemsBBLIIOGRAPHYBibliographyKaras, M., U. Bahr, F. Hillenkamp, Int. J. Mass Spectrom. Ion Proc., 1989, 92, 231–242.Nor

Página 516 - Appendix D Log Sheets

BibliographyVoyager™ Biospectrometry™ Workstation User’s Guide Bibliography-3BBLIIOGRAPHYFitzgerald, M.C., L. Zhu, and L.M. Smith, “The Analysis of Mo

Página 517

Bibliography-4 PerSeptive BiosystemsBBLIIOGRAPHYBibliographyPeptide ApplicationsBieman, K., “Mass Spectrometry of Peptides and Proteins”, Annu. Rev. B

Página 518 - Elite, linear

BibliographyVoyager™ Biospectrometry™ Workstation User’s Guide Bibliography-5BBLIIOGRAPHYTang, K., N.I. Taranenko, S.L. Allman, C.H. Chen, L.Y. Chang,

Página 519

Bibliography-6 PerSeptive BiosystemsBBLIIOGRAPHYBibliography

Página 520 - F.1 Calibration Compounds

Voyager™ Biospectrometry™ Workstation User’s Guide Glossary-1GlossaryGOSSLARYa, b, and c ions —Generic description of potential ions that are formed b

Página 521 - Workstation User’s Guide F-3

Glossary-2 PerSeptive BiosystemsGOSSLARYGlossaryDelay Time —Used in Delayed Extraction mode. Time in nanoseconds (after the laser ionizes the sample)

Página 522 - Time for Typical Standards

GlossaryVoyager™ Biospectrometry™ Workstation User’s Guide Glossary-3GOSSLARYIonization —Conversion of sample in solid, gaseous, or liquid phase to io

Página 523 - Workstation User’s Guide F-5

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-2711.5.3.2 Voyager-DE PRO Vacuum SystemVacuumc

Página 524 - Cleavages for Angiotensin

Glossary-4 PerSeptive BiosystemsGOSSLARYGlossaryMatrix —Low-mass, UV-absorbing substance used in MALDI technology to enable sample ionization. Common

Página 525 - PSD Fragments in Angiotensin

GlossaryVoyager™ Biospectrometry™ Workstation User’s Guide Glossary-5GOSSLARYTC2 (Thermocouple vacuum gauge) —Pressure gauge that operates from 1 mill

Página 526 - F-8 PerSeptive Biosystems

Glossary-6 PerSeptive BiosystemsGOSSLARYGlossary

Página 527

Voyager™ Biospectrometry™ Workstation User’s Guide Index-1IndexIDEXNNumerics384 well platePLT file 3-51position/row diagram 3-503-HPAchemical structur

Página 528 - Maintenance Log

IDEXNIndex-2 PerSeptive BiosystemsIndexAcquiring data, Instrument Control Panel (continued)Automatic Control mode 6-34Data Storage, setting 6-14data,

Página 529

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-3IDEXNAdobe Acrobat Readerinstalling from Voyager CD 2-26opening PDF Voyager files 2-26A

Página 530 - (.BIC) Setting

IDEXNIndex-4 PerSeptive BiosystemsIndexAutomatic Control mode (continued)optimizing BIC for Sequence run 6-65optimizing BIC in Manual Control mode bef

Página 531 - Workstation User’s Guide H-3

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-5IDEXNBA2, Voyager-DE STRdescription 1-40E09 error 8-25pressure range 1-41pressure, disp

Página 532 - (.BIC) Files

IDEXNIndex-6 PerSeptive BiosystemsIndexCalibration (continued)impact of changing Grid Voltage% 5-41impact of Low Mass Gate 5-79increasing accuracy 3-1

Página 533 - Optimizing

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-7IDEXNCalibration, PSDdefault 7-8default, selecting 7-26equation 7-8in Instrument Settin

Página 534 - Extraction Mode

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-28 PerSeptive Biosystems1Vacuum pumpsThree vacuum pumps create the vacuum envi

Página 535 - Continuous Extraction Mode

IDEXNIndex-8 PerSeptive BiosystemsIndexComputer (continued)technical support for altered configuration B-1troubleshooting 8-6with Signatec 500 MHz dig

Página 536 - Ratio, and Laser Threshold

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-9IDEXNCrystallization3-HPA 3-35alpha-cyano 3-34desired pattern 3-34DHB 3-35DHBs 3-35exam

Página 537 - Workstation User’s Guide H-9

IDEXNIndex-10 PerSeptive BiosystemsIndexDefault calibrationdescription 6-7equation 6-9PSD 7-8selecting 5-10Default layout, Instrument Control Panel 4-

Página 538 - H-10 PerSeptive Biosystems

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-11IDEXNDHBschemical structure and molecular weight C-9concentration 3-14crystals 3-14, 3

Página 539

IDEXNIndex-12 PerSeptive BiosystemsIndexEEjecting sample plates, Instrument Control Panel 4-25Electromagnetic compliance xxiiiEMC standards xxiiiEMIS

Página 540 - H.2.2.1 Overview

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-13IDEXNFore pump, Voyager-DE STRfunction 1-38vacuum gauge 1-40Foreline valve location 1-

Página 541

IDEXNIndex-14 PerSeptive BiosystemsIndexGuide Wire Voltage%adjusting for CID 7-16description 5-18, 5-46effect of changing 5-46effect on resolution 5-3

Página 542 - H-14 PerSeptive Biosystems

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-15IDEXNHot spots in signal intensity 6-5How to use this guide xxviiHPA, see 3-HPAHPLC-gr

Página 543

IDEXNIndex-16 PerSeptive BiosystemsIndexInstrument Control Panelbasics 4-1BIC loaded at end of Sequence run 6-79control buttons 4-24cursor and grid, d

Página 544 - H-16 PerSeptive Biosystems

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-17IDEXNInstrument Settings parameters (continued)Grid Voltage%, Continuous Extraction H-

Página 545 - H.2.3 Checking Resolution

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-2911.5.4 Vacuum Gauge PanelThe Vacuum Gauge Pa

Página 546 - H-18 PerSeptive Biosystems

IDEXNIndex-18 PerSeptive BiosystemsIndexInternal standardcalibration 6-7, 6-25concentration 3-18mass range 3-18Internal-Update calibration, see Calibr

Página 547

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-19IDEXNLaser (continued)safety information xxvstopping I-7threshold, see Laser threshold

Página 548 - H-20 PerSeptive Biosystems

IDEXNIndex-20 PerSeptive BiosystemsIndexLinking traces 4-13Live Spectrum tracedefinition 4-14during acquisition 6-16not displayed on oscilloscope syst

Página 549

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-21IDEXNManual Laser/Sample Positioning control pagedisplaying 4-27laser position 4-31loc

Página 550 - H-22 PerSeptive Biosystems

IDEXNIndex-22 PerSeptive BiosystemsIndexMatrix (continued)selecting 3-3, C-1selecting type in calibration 5-20sinapinic acid 3-3, 3-7, C-6solutions C-

Página 551

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-23IDEXNNNa adduct ioneffect on masses 8-17from buffer 3-18Nameinstrument, specifying 2-3

Página 552

IDEXNIndex-24 PerSeptive BiosystemsIndexPPage control, types of 4-9Parent ion, see PSD precursor spectrumPassword, obtaining from system administrator

Página 553

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-25IDEXNPLT files (continued)search pattern file for 3-57selecting 3-43, 3-45x,y coordina

Página 554 - Oscilloscope

IDEXNIndex-26 PerSeptive BiosystemsIndexPrinterconnecting, Voyager-DE and Voyager-DE PRO 2-18connecting, Voyager-DE STR 2-20dedicating to landscape or

Página 555 - I.1 Guidelines for Acquiring

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-27IDEXNPSD Mirror Ratio (continued)function 7-2precision displayed when you click on ent

Página 556 - I.2 Scaling

Table of Contentsvi PerSeptive BiosystemsChapter 4 Voyager Instrument Control Panel Basics4.1 Instrument Control Panel ...

Página 557 - Workstation User’s Guide I-5

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-30 PerSeptive Biosystems1• EMIS—Turns BA1 and BA2 on or off. Used during troub

Página 558 - I-6 PerSeptive Biosystems

IDEXNIndex-28 PerSeptive BiosystemsIndexPSD segments (continued)mass range 7-34number of, and composite spectrum resolution 7-6number to acquire 7-6,

Página 559 - I.3 Using the Control Stick

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-29IDEXNResolution, mass (continued)PSD mode 7-44PSD segment, optimum observed near Max S

Página 560 - I-8 PerSeptive Biosystems

IDEXNIndex-30 PerSeptive BiosystemsIndexSample concentration (continued)in IAA 3-15in sinapinic acid 3-7in THAP 3-10low concentration application tech

Página 561 - Bibliography

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-31IDEXNSample plate, types ofapplications for 3-48disposable, applications 3-49disposabl

Página 562 - MALDI Applications

IDEXNIndex-32 PerSeptive BiosystemsIndexSensitivity (continued)range, Voyager-DE STR 1-6troubleshooting 8-10Vertical Scale parameter, effect on 5-28SE

Página 563 - MS/MS Interpretation

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-33IDEXNSET file, Sequence Control Panelcreating 6-67defaults used if none specified 6-71

Página 564 - DNA Applications

IDEXNIndex-34 PerSeptive BiosystemsIndexSinapinic acidchemical structure and molecular weight C-6concentration 3-7crystals 3-7, 3-34initial velocity s

Página 565

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-35IDEXNSpectrum window (continued)Grid, displaying 4-21live data, definition 4-14Live Sp

Página 566

IDEXNIndex-36 PerSeptive BiosystemsIndexTTC2, Voyager-DEdescription 1-26max load pressure 2-32pressure range 1-30wait time 2-32TC2, Voyager-DE PROdesc

Página 567 - Glossary

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-37IDEXNTraces (continued)types of 4-14white, does not print 4-18zooming 4-13Trihydroxy a

Página 568

Parts of the Voyager-DE and Voyager-DE PRO SystemsVoyager™ Biospectrometry™ Workstation User’s Guide 1-3111.5.5 Computer ComponentsHardwareThe Voyage

Página 569

IDEXNIndex-38 PerSeptive BiosystemsIndexVacuum system, Voyager-DE PROsee also Vacuum gauge panelchambers 1-25, 1-27diagram 1-27function 1-24gauges 1-2

Página 570

IndexVoyager™ Biospectrometry™ Workstation User’s Guide Index-39IDEXNVoyager-DE PRO Biospectrometry Workstation (continued)specifications A-4startup a

Página 571

IDEXNIndex-40 PerSeptive BiosystemsIndex

Página 572

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-32 PerSeptive Biosystems11.6 Parts of the Voyager-DE STR SystemThis section d

Página 573 - Numerics

Parts of the Voyager-DE STR SystemVoyager™ Biospectrometry™ Workstation User’s Guide 1-331Parts of the Voyager-DE STR system include:• Mass spectromet

Página 574 - Index-2 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-34 PerSeptive Biosystems11.6.2 Mass SpectrometerThe parts of the Voyager-DE S

Página 575

Parts of the Voyager-DE STR SystemVoyager™ Biospectrometry™ Workstation User’s Guide 1-351Parts of the massspectrometerThe Voyager-DE STR mass spectro

Página 576 - Index-4 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-36 PerSeptive Biosystems1• Vacuum system—A pumping system and a sealed enclosu

Página 577

Parts of the Voyager-DE STR SystemVoyager™ Biospectrometry™ Workstation User’s Guide 1-371The single-stage design provides high mass resolution across

Página 578 - Index-6 PerSeptive Biosystems

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-38 PerSeptive Biosystems11.6.3 Vacuum SystemOverviewThe Voyager-DE STR Biospe

Página 579

Parts of the Voyager-DE STR SystemVoyager™ Biospectrometry™ Workstation User’s Guide 1-391Figure 1-20 Voyager-DE STR Biospectrometry Workstation Vacu

Página 580 - Index-8 PerSeptive Biosystems

Table of ContentsVoyager Biospectrometry Workstation User’s Guide vii5.2 Instrument Settings Parameter Descriptions ...

Página 581

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-40 PerSeptive Biosystems1Vacuum gaugesThe Voyager-DE STR Biospectrometry Works

Página 582

Parts of the Voyager-DE STR SystemVoyager™ Biospectrometry™ Workstation User’s Guide 1-411CAUTIONDo not press any other buttons on the panel. Pressing

Página 583 - Vertical Scale 5-28

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-42 PerSeptive Biosystems11.6.5 Front Panel IndicatorsThe front panel of the V

Página 584

Parts of the Voyager-DE STR SystemVoyager™ Biospectrometry™ Workstation User’s Guide 1-4311.6.6 Computer ComponentsHardwareThe Voyager-DE STR Biospec

Página 585

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-44 PerSeptive Biosystems1SoftwareThe Voyager-DE STR Biospectrometry Workstatio

Página 586

Software OverviewVoyager™ Biospectrometry™ Workstation User’s Guide 1-4511.7 Software OverviewThe Voyager Biospectrometry Workstation software includ

Página 587

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-46 PerSeptive Biosystems1• Ability to zoom in on up to four different areas of

Página 588

Software OverviewVoyager™ Biospectrometry™ Workstation User’s Guide 1-471Sequence ControlPanelThe Sequence Control Panel works in conjunction with the

Página 589

Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations1-48 PerSeptive Biosystems11.7.2 Post-Processing Software (Data Explorer)Data E

Página 590

2ChapterVoyager™ Biospectrometry™ Workstation User’s Guide 2-12 Installing theVoyager™Biospectrometry™WorkstationsThis chapter contains the following

Página 591

Table of Contentsviii PerSeptive BiosystemsChapter 6 Acquiring Mass Spectra6.1 Before You Begin...

Página 592

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-2 PerSeptive Biosystems22.1 Installing the SystemYour Voyager Biospectrometry

Página 593

Selecting the SiteVoyager™ Biospectrometry™ Workstation User’s Guide 2-32Allow 4 inches (10 cm) on the right side of the mass spectrometer for cables.

Página 594 - Vertical Scale 5-28, 5-50

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-4 PerSeptive Biosystems2To select input voltage:1. Remove the power cord from

Página 595

Selecting the SiteVoyager™ Biospectrometry™ Workstation User’s Guide 2-52WARNINGFIRE HAZARD. Using a fuse of the wrong type or rating can cause a fire

Página 596

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-6 PerSeptive Biosystems22.2.2 Voyager-DE STR WorkstationIn this sectionThis s

Página 597

Connecting Voyager-DE and Voyager-DE PRO WorkstationsVoyager™ Biospectrometry™ Workstation User’s Guide 2-72CAUTIONBefore operation, internal jumpers

Página 598

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-8 PerSeptive Biosystems22.3.1 Side Panel Diagrams for Mass Spectrometer and C

Página 599

Connecting Voyager-DE and Voyager-DE PRO WorkstationsVoyager™ Biospectrometry™ Workstation User’s Guide 2-92ComputerDepending on your digitizer option

Página 600

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-10 PerSeptive Biosystems2Figure 2-4 Digitizer OptionsPB100784CH1 CH2Not usedT

Página 601

Connecting Voyager-DE and Voyager-DE PRO WorkstationsVoyager™ Biospectrometry™ Workstation User’s Guide 2-1122.3.2 Connecting the Mass Spectrometer t

Página 602

Table of ContentsVoyager Biospectrometry Workstation User’s Guide ix6.6.6 Process that Occurs when Accumulating Spectra from Multiple Search Pattern

Página 603

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-12 PerSeptive Biosystems22.3.3 Connecting the Signatec 500 MHz Digitizer Boar

Página 604

Connecting Voyager-DE and Voyager-DE PRO WorkstationsVoyager™ Biospectrometry™ Workstation User’s Guide 2-1322.3.4 Connecting the LSA1000 LeCroy Digi

Página 605

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-14 PerSeptive Biosystems2Refer to the following table when connecting the LSA1

Página 606

Connecting Voyager-DE and Voyager-DE PRO WorkstationsVoyager™ Biospectrometry™ Workstation User’s Guide 2-1522.3.5 Connecting the Tektronix Oscillosc

Página 607

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-16 PerSeptive Biosystems2Refer to the following table when connecting the osci

Página 608

Connecting Voyager-DE and Voyager-DE PRO WorkstationsVoyager™ Biospectrometry™ Workstation User’s Guide 2-1722.3.6 Connecting the Video Monitor Figur

Página 609

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-18 PerSeptive Biosystems2Refer to the following table when connecting the vide

Página 610

Connecting the Voyager-DE STR WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 2-1922.4 Connecting the Voyager-DE STR WorkstationFigure

Página 611

Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations2-20 PerSeptive Biosystems2Figure 2-10 Computer Connections for Voyager-DE STR

Página 612

Connecting the Voyager-DE STR WorkstationVoyager™ Biospectrometry™ Workstation User’s Guide 2-212NOTE: The computer layout may change without notice.

Comentários a estes Manuais

Sem comentários