Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-46 PerSeptive Biosystems
2
2.8 Startup and Shutdown
In this section
This section describes:
• Powering up
•Initializing
• Reinitializing
• Powering down system components
• Powering down the mass spectrometer
Powering up
To power up the Voyager Workstation:
1. Turn on the main power switch. The power switch is
located:
• On the right side panel of the mass spectrometer
cabinet on Voyager-DE and Voyager-DE PRO
systems.
• On the back panel of the mass spectrometer
cabinet on the Voyager-DE STR system.
The vacuum pumping system begins running.
After two minutes, the turbo pump starts. You will hear a
whine as the vacuum system spins up to speed.
2. Turn on remaining system devices in this order:
• Video monitor
• External digitizer, if your system includes one
• Oscilloscope, if your system includes one
• Computer
• Printer
3. If your system includes an oscilloscope, observe the
oscilloscope and wait approximately one minute until it
completes its initialization cycle. When the cycle is
complete, a message stating that the power on self check
passed is displayed, then the oscilloscope screen is
displayed. See Appendix I, Using the Oscilloscope and
Control Stick, for more information.