Evaluating Data in the Instrument Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-31
6
Figure 6-13 Resolution Calculator Results
The table below lists a general rating scale for resolution and
molecular weight ranges for compounds acquired in Linear
mode.
Resolution
result
Table 6-2 Resolution Rating Scale
Compounds <
2,000 Da
Compounds 2,000
to 5,000 Da
Compounds 5,000
to 25,000 Da
Compounds >
25,000 Da
Resolution Rating Resolution Rating Resolution Rating Resolution Rating
500 Not
acceptable
250 Not
acceptable
500 Not
acceptable
50
Acceptable
1,500 Acceptable 400 Acceptable 700 Acceptable 100
Good
2,000 Good 500 Good 900 Good >200
Very good
2,500 Very good >600 Very good 1,000 Very good